A new LCR meter right angle phase self-calibration method for use in rf resistance calibrations by using capacitance standards

机译:一种新的LCR表直角相位自校准方法,该方法用于通过使用电容标准进行RF电阻校准

获取原文

摘要

This paper describes how to calibrate the right angle phase of a Four-Terminal-Pair (4TP) LCR meter. The meter uses the right angle signal to process synchronous detector as a vector receiver. The right angle calibration method uses, at least, one programmable phase signal generator and a phase shift circuits. The signal generator and the phase shifter do not require a precision 90 degree angle. As a result, the right angle phase deviation of the LCR meter is better than 100 murad at 10 MHz. This method is also applicable to the phase calibration of a vector network analyzer.
机译:本文介绍了如何校准四端对(4TP)LCR计的直角相位。该仪表使用直角信号来处理同步检测器作为矢量接收器。直角校准方法至少使用一个可编程相位信号发生器和相移电路。信号发生器和移相器不需要精度90度角。结果,LCR仪表的直角相位偏差在10MHz的10 MHz中优于100村。该方法也适用于矢量网络分析仪的相位校准。