Absolute accuracy of the line-focus-beam (LFB) acoustic microscopy system is investigated for measurements of the leaky surface acoustic wave (LSAW) velocity and attenuation, and a method of system calibration is proposed. In order to discuss the accuracy, it is necessary to introduce a standard specimen whose bulk acoustic properties, (e.g., the independent elastic constants and density) are measured with high accuracy. Single crystal substrates of gadolinium gallium garnet (GGG) are taken as standard specimens. The LSAW propagation characteristics are measured and compared with the calculated results using the measured bulk acoustic properties. Calibration is demonstrated for the system using two LFB acoustic lens devices with a cylindrical concave surface of 1-mm radius in the frequency range 100 to 300 MHz.
机译:研究了线聚焦束(LFB)声学显微镜系统用于测量泄漏声表面波(LSAW)速度和衰减的绝对精度,并提出了一种系统校准的方法。为了讨论精度,有必要引入一种标准试样,该试样的体声特性(例如,独立的弹性常数和密度)可以高精度地测量。 g镓石榴石(GGG)的单晶衬底被用作标准样品。测量LSAW传播特性,并使用测得的体声特性与计算结果进行比较。演示了使用两个LFB声透镜设备对系统进行校准的情况,该设备在100至300 MHz的频率范围内具有1mm半径的圆柱形凹面。