首页> 外文期刊>IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control >Noninvasive transesophageal cardiac thermal ablation using a 2-D focused, ultrasound phased array: a simulation study
【24h】

Noninvasive transesophageal cardiac thermal ablation using a 2-D focused, ultrasound phased array: a simulation study

机译:使用二维聚焦超声相控阵无创经食道心脏热消融术的模拟研究

获取原文
获取原文并翻译 | 示例
           

摘要

This simulation study proposes a noninvasive, transesophageal cardiac-thermal ablation using a planar ultrasound phased array (1 MHz, 60 /spl times/ 10 mm/sup 2/, 0.525 mm interelement spacing, 114 /spl times/ 20 elements). Thirty-nine foci in cardiac muscle were defined at 20, 40, and 60-mm distances and at various angles from the transducer surface to simulate the accessible posterior left atrial wall through the esophageal wall window. The ultrasound pressure distribution and the resulting thermal effect in a volume of 60 /spl times/ 80 /spl times/ 80 mm/sup 3/, including esophagus and cardiac muscle, were simulated for each focus. For 1, 10, and 20-s sonications with 60/spl deg/C and 70/spl deg/C peak temperatures in cardiac muscle and without thermal damage in esophageal wall, the transducer acoustic powers were 105-727, 28-117, 21-79 W and 151-1044, 40-167, 30-114 W, respectively. The simulated lesions (thermal dose in equivalent minutes at 43/spl deg/C /spl ges/240 minutes) at these foci had lengths of 1-6, 3-11, 3-13 mm and 3-15, 5-19, 6-23 mm, respectively, and widths of 1-4, 2-7, 3-9 mm and 3-9, 4-13, 4-17 mm, respectively. As a first step toward feasibility, controllable tissue coagulation in cardiac tissue without damage to the esophagus was demonstrated numerically.
机译:该模拟研究提出了一种使用平面超声相控阵(1 MHz,60 / spl次/ 10 mm / sup 2 /,0.525 mm元素间距,114 / spl次/ 20个元素)的无创经食道心脏热消融术。在距换能器表面20毫米,40毫米和60毫米的距离和各个角度处定义了心肌中的39个病灶,以模拟通过食道壁窗可触及的左后房壁。针对每个焦点,模拟了超声压力分布和所产生的热效应,其体积为60 / spl次/ 80 / spl次/ 80 mm / sup 3 /,包括食道和心肌。对于1s,10s和20s超声,心肌中的峰值温度为60 / spl deg / C和70 / spl deg / C,并且食管壁没有热损伤,换能器的声功率为105-727、28-117, 21-79 W和151-1044、40-167、30-114W。在这些病灶处的模拟病变(在43 / spl deg / C / spl ges / 240分钟的等效分钟内的热剂量)的长度为1-6、3-11、3-13毫米和3-15、5-19,分别为6-23毫米和1-4、2-7、3-9毫米和3-9、4-13、4-17毫米的宽度。作为迈向可行性的第一步,在数值上证明了可控制的心脏组织凝结过程不会损坏食道。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号