Adaptive Clutter Demodulation for Non-Contrast Ultrasound Perfusion Imaging

机译:自适应杂波解调,用于非造影超声灌注成像

获取原文
获取原文并翻译 | 示例

摘要

Conventional Doppler ultrasound is useful for visualizing fast blood flow in large resolvable vessels. However, frame rate and tissue clutter caused by movement of the patient or sonographer make visualizing slow flow with ultrasound difficult. Patient and sonographer motion causes spectral broadening of the clutter signal, which limits ultrasound's sensitivity to velocities greater than 5-10 mm/s for typical clinical imaging frequencies. To address this, we propose a clutter filtering technique that may increase the sensitivity of Doppler measurements to at least as low as 0.52 mm/s. The proposed technique uses plane wave imaging and an adaptive frequency and amplitude demodulation scheme to decrease the bandwidth of tissue clutter. To test the performance of the adaptive demodulation method at suppressing tissue clutter bandwidths due to sonographer hand motion alone, six volunteer subjects acquired data from a stationary phantom. Additionally, to test in vivo feasibility, arterial occlusion and muscle contraction studies were performed to assess the efficiency of the proposed filter at preserving signals from blood velocities 2 mm/s or greater at a 7.8 MHz center frequency. The hand motion study resulted in initial average bandwidths of 175 Hz (8.60 mm/s), which were decreased to 10.5 Hz (0.52 mm/s) at -60 dB using our approach. The in vivo power Doppler studies resulted in 4.73 dB and 4.80 dB dynamic ranges of the blood flow with the proposed filter and 0.15 dB and 0.16 dB dynamic ranges of the blood flow with a conventional 50 Hz high-pass filter for the occlusion and contraction studies, respectively.
机译:常规多普勒超声可用于观察大型可分辨血管中的快速血流。但是,由于患者或超声医师的运动引起的帧频和组织混乱,很难用超声波观察缓慢的血流。患者和超声医师的运动会导致杂波信号的频谱展宽,这限制了超声对典型临床成像频率大于5-10 mm / s的速度的敏感性。为了解决这个问题,我们提出了一种杂波滤波技术,可以将多普勒测量的灵敏度提高到至少低至0.52 mm / s。所提出的技术使用平面波成像和自适应频率和幅度解调方案来减少组织杂波的带宽。为了测试自适应解调方法在抑制仅由于超声医师的手运动引起的组织混乱带宽方面的性能,六名志愿者受试者从固定体模中获取了数据。另外,为了测试体内可行性,进行了动脉闭塞和肌肉收缩研究,以评估所提出的过滤器在保持中心频率为7.8 MHz时血流速度为2 mm / s或更高的信号时的效率。手动研究得出的初始平均带宽为175 Hz(8.60 mm / s),使用我们的方法在-60 dB时降低至10.5 Hz(0.52 mm / s)。体内功率多普勒研究的结果是,使用拟议的过滤器,血流的动态范围为4.73 dB和4.80 dB;使用常规50 Hz高通滤波器的阻塞和收缩研究,血流的动态范围为0.15 dB和0.16 dB。 , 分别。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号