Biosafety Assessment and Airborne Fungal Concentration as an Evaluation Index in University Laboratories and Hospital Diagnostic Korean Laboratories

机译:生物安全性评估和空气中真菌的浓度作为大学实验室和医院诊断韩国实验室的评估指标

获取原文
获取原文并翻译 | 示例
           

摘要

This study evaluated both biosafety levels (BLs) and airborne fungal concentrations in microbiological laboratories in Seoul, Korea. To evaluate biosafety facilities, we used a checklist containing 67 questions in nine categories. We also measured airborne fungal concentration according to the BLs. Airborne fungal concentrations were higher in BL-1 facilities (240 CFU/m~3) than in BL-2 facilities (25 CFU/m~3). The airborne fungal concentrations significantly differed among the laboratories that were graded as poor, fair, and good. Especially, a significant negative correlation was observed between the airborne fungal concentrations and the biosafety levels (p = .001). We recommend that the guidelines of biosafety be followed to improve laboratory environment.
机译:这项研究评估了韩国首尔市微生物实验室的生物安全水平(BLs)和空气传播的真菌浓度。为了评估生物安全设施,我们使用了一个清单,其中包含9个类别的67个问题。我们还根据BLs测量了空气传播的真菌浓度。 BL-1设施(240 CFU / m〜3)中的空气传播真菌浓度高于BL-2设施(25 CFU / m〜3)。在空气实验室中,空气传播的真菌浓度差异显着,被分为差,中等和良好。尤其是,在空气中传播的真菌浓度与生物安全水平之间观察到显着的负相关性(p = .001)。我们建议遵循生物安全准则以改善实验室环境。