首页> 外文期刊>Human and ecological risk assessment >Levels of Total Airborne Bacteria, Gram-Negative Bacteria, and Endotoxin According to Biosafety Levels in Korean Biosafety Laboratories
【24h】

Levels of Total Airborne Bacteria, Gram-Negative Bacteria, and Endotoxin According to Biosafety Levels in Korean Biosafety Laboratories

机译:根据韩国生物安全实验室的生物安全水平,确定总的空气传播细菌,革兰氏阴性细菌和内毒素水平

获取原文
获取原文并翻译 | 示例
           

摘要

We assessed in this study the total airborne bacteria (TAB), gram-negative bacteria (GNB), and endotoxin (ET) concentrations in three Korean university laboratories and two hospital diagnostic laboratories categorized as biosafety level 1 (BL 1) or biosafety level 2 (BL 2). We also investigated the concentrations of TAB, GNB, and ET relative to the performance results of biosafety cabinets (BSCs). TAB and GNB were incubated, and ET was analyzed using the kinetic Limulus amebocyte lysate assay. A total of 246 (TAB = 95, GNB = 95, ET = 56) air samples were collected from nine laboratories. TAB, GNB, and ET concentrations were significantly higher in BL 1 laboratories and in laboratories that failed the BSC performance test compared to BL 2 laboratories and laboratories that passed the BSC performance test. Correlation coefficients were calculated to determine the relationships between TAB, GNB, and ET, and it was found that TAB significantly correlated with GNB (r = 0.35, p < .01) and ET (r = 0.29, p < .01). Careful attention needs to be paid to BL 1 laboratories, and BSCs should be managed on a regular basis to improve the air quality of university and hospital laboratories.
机译:我们在这项研究中评估了三个韩国大学实验室和两个医院诊断实验室中总空气传播细菌(TAB),革兰氏阴性细菌(GNB)和内毒素(ET)的浓度,这些实验室被归类为生物安全等级1(BL 1)或生物安全等级2 (BL 2)。我们还研究了相对于生物安全柜(BSC)性能结果的TAB,GNB和ET的浓度。温育TAB和GNB,并使用动力Li变形细胞溶解物测定法分析ET。从9个实验室收集了246个空气样品(TAB = 95,GNB = 95,ET = 56)。与BL 2实验室和通过BSC性能测试的实验室相比,BL 1实验室和未通过BSC性能测试的实验室的TAB,GNB和ET浓度显着更高。计算相关系数以确定TAB,GNB和ET之间的关系,发现TAB与GNB(r = 0.35,p <.01)和ET(r = 0.29,p <.01)显着相关。需要特别注意BL 1实验室,并且应定期管理BSC,以改善大学和医院实验室的空气质量。

著录项

  • 来源
    《Human and ecological risk assessment》 |2013年第6期|1576-1585|共10页
  • 作者单位

    Institute of Health and Environment, School of Public Health, Seoul National University, Seoul, South Korea;

    Department of Chemical Engineering and Chemistry, Inha University, Incheon, South Korea;

    CIHm, Occupational & Environmental Health Lab, Department of Environmental Health, School of Public Health, Seoul National Univer-sity, Gwanak 599, Gwanak-ro, Seoul 151-742, South Korea;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    airborne bacteria; biosafety cabinet; endotoxin; gram-negative bacteria;

    机译:空气传播的细菌;生物安全柜;内毒素革兰氏阴性菌;