首页> 外文期刊>European Journal of Applied Physiology >Comparison of three mobile devices for measuring R–R intervals and heart rate variability: Polar S810i, Suunto t6 and an ambulatory ECG system
【24h】

Comparison of three mobile devices for measuring R–R intervals and heart rate variability: Polar S810i, Suunto t6 and an ambulatory ECG system

机译:三种测量R–R间隔和心率变异性的移动设备的比较:Polar S810i,Suunto t6和动态ECG系统

获取原文
获取原文并翻译 | 示例
           

摘要

The first aim of this study was to compare an ambulatory five-lead ECG system with the commercially available breast belt measuring devices; Polar S810i and Suunto t6, in terms of R–R interval measures and heart rate variability (HRV) indices. The second aim was to compare different HRV spectral analysis methods. Nineteen young males (aged between 22 and 31 years, median 24 years) underwent simultaneous R–R interval recordings with the three instruments during supine and sitting rest, moderate dynamic, and moderate to vigorous static exercise of the upper and lower limb. For each subject, 17 R–R interval series of 3-min length were extracted from the whole recordings and then analyzed in frequency domain using (1) a fast Fourier transform (FFT), (2) an autoregressive model (AR), (3) a Welch periodogram (WP) and (4) a continuous wavelet transform (CWT). Intra-class correlation coefficients (ICC) and Bland–Altman limits of agreement (LoA) method served as criteria for measurement agreement. Regarding the R–R interval recordings, ICC (lower ICC 95% confidence interval >0.99) as well as LoA (maximum LoA: −15.1 to 14.3 ms for ECG vs. Polar) showed an excellent agreement between all devices. Therefore, the three instruments may be used interchangeably in recording and interpolation of R–R intervals. ICCs for HRV frequency parameters were also high, but in most cases LoA analysis revealed unacceptable discrepancies between the instruments. The agreement among the different frequency transform methods can be taken for granted when analyzing the normalized power in low and high frequency ranges; however, not when analyzing the absolute values.
机译:这项研究的首要目的是将可移动的五导联心电图系统与市售的胸带测量设备进行比较。根据R–R间隔测量和心率变异性(HRV)指标,Polar S810i和Suunto t6。第二个目的是比较不同的HRV光谱分析方法。 19名年轻男性(年龄在22至31岁之间,中位年龄为24岁)在仰卧和坐卧,上肢和下肢中等程度的运动以及中度到剧烈的静态运动期间同时使用这三种仪器进行了R–R间隔记录。对于每个主题,从整个记录中提取了17个3分钟长度的R–R间隔序列,然后在频域中使用(1)快速傅里叶变换(FFT),(2)自回归模型(AR),( 3)韦尔奇周期图(WP)和(4)连续小波变换(CWT)。类内相关系数(ICC)和Bland–Altman一致性极限(LoA)方法用作衡量一致性的标准。关于R–R间隔记录,ICC(较低的ICC 95%置信区间> 0.99)以及LoA(最大LoA:ECG与Polar的最大LoA:-15.1到14.3 ms)显示了所有器件之间的出色一致性。因此,这三种乐器可以在记录和插补R–R间隔时互换使用。用于HRV频率参数的ICC也很高,但是在大多数情况下,LoA分析显示仪器之间的差异是无法接受的。分析低频和高频范围内的归一化功率时,可以认为不同频率变换方法之间的一致性是理所当然的。但是,在分析绝对值时不行。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号