Simultaneous Determination of 34 Amino Acids in Tumor Tissues from Colorectal Cancer Patients Based on the Targeted UHPLC-MS/MS Method

机译:基于靶向UHPLC-MS / MS法同时测定肿瘤组织中34例肿瘤组织氨基酸

获取原文
   

获取外文期刊封面封底 >>

       

摘要

A targeted ultrahigh-performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method was established and validated for the simultaneous determination of 34 amino acids in tissue samples from colorectal cancer (CRC) patients. The chromatographic separation was achieved on an Agilent ZORBAX SB-C_(18) column (3.0?×?150?mm, 5? μ m) with a binary gradient elution system (A, 0.02% heptafluorobutyric acid and 0.2% formic acid in water, v/v; B, methanol). The run time was 10?min. The multiple reaction monitoring mode was chosen with an electrospray ionization source operating in the positive ionization mode for data acquisition. The linear correlation coefficients were >0.99 for all the analytes in their corresponding calibration ranges. The sample was pretreated based on tissue homogenate and protein precipitation with a 100?mg aliquot sample. The average recovery and matrix effect for 34 amino acids and 3 internal standards were 39.00%~146.95% and 49.45%~173.63%, respectively. The intra- and interday accuracy for all the analytes ranged from ?13.52% to 14.21% (RSD ≤8.57%) and from ?14.52% to 12.59% (RSD ≤10.31%), respectively. Deviations of stability under different conditions were within ±15% for all the analytes. This method was applied to simultaneous quantification of 34 amino acids in tissue samples from 94 CRC patients.
机译:建立了靶向超高效液相色谱 - 串联质谱(UHPLC-MS / MS)方法,并验证了与结肠直肠癌(CRC)患者组织样品中的34个氨基酸的同时测定。在Agilent Zorbax SB-C_(18)柱上(3.0××150×150毫米,5Ω·μm)在水中进行色谱分离,其中二进制梯度洗脱系统(a,0.02%庚二丁酸和0.2%甲酸在水中,v / v; b,甲醇)。运行时间为10?分钟。选择多次反应监测模式,用在正电离模式下操作的电喷雾电离源进行数据采集。对于其相应的校准范围内的所有分析物,线性相关系数> 0.99。基于组织匀浆和蛋白质沉淀,用100μmg等分样品样品进行预处理样品。 34个氨基酸和3个内标的平均回收和基质效应分别为39.00%〜146.95%和49.45%〜173.63%。所有分析物的内部和间隔的准确性范围为13.52%至14.21%(RSD≤8.57%),分别为14.52%至12.59%(RSD≤10.31%)。所有分析物的不同条件下稳定性的偏差在±15%以内。将该方法应用于来自94例CRC患者的组织样品中34个氨基酸的同时定量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号