首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >The "eyetone" blood glucose reflectance colorimeter evaluated for in vitro and in vivo accuracy and clinical efficacy.
【24h】

The "eyetone" blood glucose reflectance colorimeter evaluated for in vitro and in vivo accuracy and clinical efficacy.

机译:评价“酮”血糖反射比色计的体外和体内准确性和临床疗效。

获取原文

摘要

We evaluated the performance of a blood glucose reflectance colorimeter ("Eyetone," Ames Co.) for accuracy and precision with use of "Dextrostix" (Ames Co.) glucose oxidase reagent strips for blood samples with known and unknown concentrations of glucose covering the usual range of neonatal blood glucose (200-800 mg/L). The meter was calibrated and tested by research nurses and one clinical chemist. Five unknowns were tested for accuracy and precision (56-92 determinations per unknown) and compared with Beckman Astra values (plasma and calculated whole blood). Eyetone/Dextrostix values differed (gave lower values) from the calculated whole-blood values only at concentrations less than 300 mg/L. On 258 clinical specimens from newborn infants, Eyetone/Dextrostix values were not different from calculated whole-blood values (p less than 0.05, r = 0.80). Operator training to develop a consistent procedure was the most critical factor in achieving accurate and precise results.
机译:我们使用“ Dextrostix”(Ames Co.)葡萄糖氧化酶试剂条对已知和未知浓度的葡萄糖进行了评估,评估了血糖反射比色计(“ Eyetone”,Ames Co.)的准确性和精密度。新生儿血糖的通常范围(200-800 mg / L)。该仪表由研究护士和一名临床化学家进行校准和测试。测试了五个未知物的准确性和精密度(每个未知物56-92个测定值),并与Beckman Astra值(血浆和计算出的全血)进行了比较。仅在浓度低于300 mg / L时,眼音/排毒值才与计算的全血值有所不同(降低值)。在258例来自新生儿的临床标本中,Eyetone / Dextrostix值与计算的全血值无差异(p小于0.05,r = 0.80)。对操作员进行培训以建立一致的程序是获得准确,准确结果的最关键因素。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号