We describe a procedure for the rapid and specific measurement of creatinine, in which it is separated from other compounds in serum or urine by paired-ion chromatography and is quantified by measuring its absorbance at 200 nm. The procedure can be done on as little as 10 microliter of serum. Between-day precision studies for concentrations of 13 and 62 mg/liter yielded coefficients of variation of 6.9 and 2.2%, respectively. Analytical recovery of various amounts of creatinine added to plasma exceeded 95% in all cases. The proposed procedure was compared with the continuous-flow procedure by analyzing a series of serum and urine specimens by both methods. There was excellent agreement for urine specimens, but with serum the results by the present method were significantly (P less than 0.001) lower.
机译:我们描述了一种快速,特异性地测定肌酐的方法,其中通过配对离子色谱法将肌酐与血清或尿液中的其他化合物分离,并通过测量其在200 nm处的吸光度进行定量。该过程可在低至10微升的血清上完成。对于浓度为13和62 mg / L的日间精密度研究,得出的变异系数分别为6.9和2.2%。在所有情况下,添加到血浆中的各种肌酐的分析回收率均超过95%。通过两种方法分析一系列血清和尿液标本,将拟议的程序与连续流程序进行了比较。尿液标本有很好的一致性,但采用血清方法,本方法的结果显着降低(P小于0.001)。