Pleus et al1 have shown in a recent issue of Journal Diabetes Sciences and Technology that patients seemed to spend considerably less time in hypoglycemia in two studies which used real-time blood glucose monitoring (RTBGM)2,3 than in one study which used interstitial continuous glucose monitoring (ICGM),4 that is FreeStyle Libre CGM (FSLCGM). They suggested that ICGM or RTBGM data should not be reported without accompanying BGM and that quantitative analysis of the bias between ICGM and BGM should be done. We have already reported and quantified, using FSL-CGM, the bias between ICGM and BGM data in insulin-treated dependent elderly people with diabetes living in a long-term care home. Out of 144 interstitial glucose (IG) lower than 70 mg/dL, 74 (51.4%) were associated with blood glucose (BG) equal or higher to 70 mg/dL.
机译:Pleus等[1]在最新一期的《糖尿病科学与技术杂志》上显示,与使用间质性连续监测的一项研究相比,两项使用实时血糖监测(RTBGM)2,3的患者在低血糖方面的花费似乎要少得多。血糖监测(ICGM),4,即FreeStyle Libre CGM(FSLCGM)。他们建议,ICGM或RTBGM数据应在没有随附BGM的情况下报告,而应对ICGM和BGM之间的偏差进行定量分析。我们已经使用FSL-CGM报告并量化了住在长期护理院中的接受胰岛素治疗的糖尿病依赖性老年人的ICGM和BGM数据之间的偏差。在144个低于70 mg / dL的间质葡萄糖(IG)中,有74个(51.4%)与等于或高于70 mg / dL的血糖(BG)相关。