Blood Glucose Monitoring Data Should Be Reported in Detail When Studies About Efficacy of Continuous Glucose Monitoring Systems Are Published

机译:发表有关连续血糖监测系统功效的研究时,应详细报告血糖监测数据

获取原文
           

摘要

Recently, two clinical trials about a “sensor-based flash glucose monitoring system” and its efficacy in reducing time in hypoglycemia were published. Interestingly, patients spent more time at low glucose concentrations in these studies than in other studies related to the efficacy of real-time continuous glucose monitoring (rtCGM). Although it is possible that the study populations differed from those in other studies, another potential explanation is that the CGM system used in these two studies had a negative glucose measurement bias. Such a negative bias was reported in recent literature, suggesting that the CGM system may inaccurately indicate hypoglycemia. Reporting blood glucose monitoring data would help to interpret the CGM data at least in the context of time spent in various glucose ranges as a parameter with which quality of diabetes therapy is measured.
机译:最近,发表了两项有关“基于传感器的快速血糖监测系统”及其减少低血糖时间的功效的临床试验。有趣的是,与其他与实时连续葡萄糖监测(rtCGM)功效相关的研究相比,在这些研究中患者在低葡萄糖浓度下花费了更多的时间。尽管研究人群可能与其他研究人群不同,但另一个潜在的解释是这两项研究中使用的CGM系统的葡萄糖测量偏差为负。在最近的文献中报道了这种负偏倚,表明CGM系统可能不准确地指示低血糖。报告血糖监测数据将至少在各种血糖范围内花费的时间作为测量糖尿病治疗质量的参数时,有助于解释CGM数据。