Real-time cardiovascular magnetic resonance at 1.5?T using balanced SSFP and 40?ms resolution

机译:使用平衡的SSFP和40µms分辨率在1.5?T进行实时心血管磁共振

获取原文
   

获取外文期刊封面封底 >>

       

摘要

BackgroundWhile cardiovascular magnetic resonance (CMR) commonly employs ECG-synchronized cine acquisitions with balanced steady-state free precession (SSFP) contrast at 1.5?T, recent developments at 3?T demonstrate significant potential for T1-weighted real-time imaging at high spatiotemporal resolution using undersampled radial FLASH. The purpose of this work was to combine both ideas and to evaluate a corresponding real-time CMR method at 1.5?T with SSFP contrast.MethodsRadial gradient-echo sequences with fully balanced gradients and at least 15-fold undersampling were implemented on two CMR systems with different gradient performance. Image reconstruction by regularized nonlinear inversion (NLINV) was performed offline and resulted in real-time SSFP CMR images at a nominal resolution of 1.8?mm and with acquisition times of 40?ms.ResultsStudies of healthy subjects demonstrated technical feasibility in terms of robustness and general image quality. Clinical applicability with access to quantitative evaluations (e.g., ejection fraction) was confirmed by preliminary applications to 27 patients with typical indications for CMR including arrhythmias and abnormal wall motion. Real-time image quality was slightly lower than for cine SSFP recordings, but considered diagnostic in all cases.ConclusionsExtending conventional cine approaches, real-time radial SSFP CMR with NLINV reconstruction provides access to individual cardiac cycles and allows for studies of patients with irregular heartbeat.
机译:背景技术虽然心血管磁共振(CMR)通常采用心电图同步的电影采集,并且在1.5?T时具有平衡的稳态自由进动(SSFP)对比,但在3?T时的最新进展显示出在高时空下T1加权实时成像的巨大潜力使用欠采样的径向FLASH分辨率。这项工作的目的是结合这两种思路并评估在1.5?T时具有SSFP对比度的实时CMR方法。方法在两个CMR系统上实现了具有完全平衡梯度和至少15倍欠采样的径向梯度-回波序列具有不同的渐变效果。离线执行正则化非线性反演(NLINV)进行的图像重建,得到实时SSFP CMR图像,其标称分辨率为1.8?mm,采集时间为40?ms。结果对健康受试者的研究从鲁棒性和稳定性方面证明了技术可行性。总体图像质量。通过初步应用到27例具有典型CMR指征(包括心律不齐和异常室壁运动)的患者,证实了其可用于定量评估(例如射血分数)的临床适用性。实时图像质量略低于电影SSFP记录,但在所有情况下均被认为具有诊断性。结论在传统电影方法的基础上,采用NLINV重建的实时放射状SSFP CMR可以访问各个心动周期,并允许研究心律不齐的患者。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号