Dosimetric verification and quality assurance of running‐start‐stop (RSS) delivery in tomotherapy

机译:放射疗法中的启停(RSS)输送的剂量学验证和质量保证

获取原文
           

摘要

The purpose of this study was to evaluate the dosimetric profiles and delivery accuracy of running-start-stop (RSS) delivery in tomotherapy and to present initial quality assurance (QA) results on the accuracy of the dynamic jaw motion, dosimetric penumbrae of the RSS dynamic jaw and the static jaw were measured by radiographic films. Delivery accuracy of the RSS was evaluated by gamma analysis on film measurements of 12 phantom plans. Consistency in the performance of RSS was evaluated by QA procedures over the first nine months after the installation of the feature. These QA were devised to check: 1) positional accuracy of moving jaws; 2) consistency of relative radiation output collimated by discrete and continuously sweeping jaws; 3) consistency of field widths and profiles. In the longitudinal direction, the dose penumbra in RSS delivery was reduced from 17.3 mm to 10.2 mm for 2.5 cm jaw, and from 33.2 mm to 9.6 mm for 5 cm jaw. Gamma analysis on the twelve plans revealed that over 90% of the voxels in the proximity of the penumbra region satisfied the gamma criteria of 2% dose difference and 2 mm distance-to-agreement. The initial QA results during the first nine months after installation of the RSS are presented. Jaw motion was shown to be accurate with maximum encoder error less than 0.42 mm. The consistency of relative output for discrete and continuously sweeping jaws was within 1.2%. Longitudinal radiation profiles agreed to the reference profile with maximum gamma 1 and field width error 1.8 % . With the same jaw width, RSS showed better dose penumbrae compared to those from static jaw delivery. The initial QA results on the accuracy of moving jaws, reproducibility of dosimetric output and profiles were satisfactory.PACS number: 87.55.km
机译:这项研究的目的是评估在放射疗法中剂量分布图和运行开始-停止(RSS)递送的递送准确性,并就动态颌骨运动,RSS的剂量半影的准确性提出初步质量保证(QA)结果动态颌骨和静态颌骨用射线照相胶片测量。通过对12个幻像计划的胶片测量值进行伽玛分析来评估RSS的传送准确性。安装此功能后的前9个月,通过质量检查程序评估了RSS性能的一致性。这些质量检查旨在检查:1)活动钳口的位置精度; 2)通过离散和连续扫动的钳口准直相对辐射输出的一致性; 3)字段宽度和轮廓的一致性。在纵向方向上,RSS输送的半影剂量从2.5 cm的下颚从17.3 mm减小到10.2 mm,而5 cm的下颚从33.2 mm减小到9.6 mm。对这十二个计划的伽玛分析显示,半影区附近超过90%的体素满足伽玛标准,即2%的剂量差异和2毫米的协议距离。介绍了安装RSS后的前九个月的初始质量检查结果。颚夹运动被证明是准确的,最大编码器误差小于0.42 mm。离散和连续清扫钳口的相对输出一致性在1.2%以内。纵向辐射剖面与参考剖面一致,最大伽玛值为1,场宽误差为1.8%。在相同的颌骨宽度下,与静态颌骨递送相比,RSS显示出更好的剂量半影。最初的质量检查结果表明活动钳口的准确性,剂量输出的可重复性和轮廓令人满意.PACS数:87.55.km

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号