Reproducibility of the output signal in Differential Scanning Calorimeter has been examined in detail. The ordinate calibration factor has been found to diminish at higher temperatures. Calibration factor for enthalpy output has also been evaluated. A procedure for calibrating the area by the proper selection of the base-line for both scanning and isothermal modes of operation has been described. This factor has largely been ignored in the papers published so far. And also the assessment of the performance of Thermomechanical analyser has been done by using AI as expansion standard varying heating rate, sensitivity and load. A simple method for the determination Youngs modulus at room temperature using TMA is also given.
机译:已经详细检查了差示扫描量热仪中输出信号的可重复性。已经发现纵坐标校准因子在较高温度下会减小。焓输出的校准系数也已评估。一种用于通过碱基线路用于扫描和操作的等温模式的适当选择校准区域过程进行了说明。迄今为止,该因素在很大程度上已被忽略。此外,还通过使用AI作为改变加热速率,灵敏度和负载的扩展标准,对热机械分析仪的性能进行了评估。还给出了一种使用TMA测定室温下杨氏模量的简单方法。