首页> 外文期刊>Proceedings >In-Vivo Implantable Sensor System for Measuring Bladder Wall Movements
【24h】

In-Vivo Implantable Sensor System for Measuring Bladder Wall Movements

机译:用于测量膀胱壁运动的体内植入式传感器系统

获取原文
           

摘要

The design and implementation of a medical implantable device for bladder monitoring is presented. A network of accelerometers is to be implanted in-between the outside bladder muscle and the mucous membrane. A pressure sensor is also implanted in this submucosal layer to obtain intraluminal pressure readings. The sensor system is powered by an implanted battery and provides a wireless communication link for data and control. A smart measurement protocol allows the system to be operational for several weeks. The sensors are mounted on a flexible printed circuit board that can be elongated up to 250%, to accommodate the natural stretching of the bladder organ tissue during filling.
机译:介绍了用于膀胱监测的医疗植入式设备的设计和实现。加速度计的网络将植入膀胱外肌和粘膜之间。压力传感器也被植入该粘膜下层以获得腔内压力读数。传感器系统由植入式电池供电,并提供用于数据和控制的无线通信链接。智能测量协议可使系统运行数周。传感器安装在可延长至250%的柔性印刷电路板上,以适应填充过程中膀胱器官组织的自然拉伸。