首页> 外文OA文献 >In-Vivo Implantable Sensor System for Measuring Bladder Wall Movements
【2h】

In-Vivo Implantable Sensor System for Measuring Bladder Wall Movements

机译:用于测量膀胱壁运动的活体内植入式传感器系统

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The design and implementation of a medical implantable device for bladder monitoring is presented. A network of accelerometers is to be implanted in-between the outside bladder muscle and the mucous membrane. A pressure sensor is also implanted in this submucosal layer to obtain intraluminal pressure readings. The sensor system is powered by an implanted battery and provides a wireless communication link for data and control. A smart measurement protocol allows the system to be operational for several weeks. The sensors are mounted on a flexible printed circuit board that can be elongated up to 250%, to accommodate the natural stretching of the bladder organ tissue during filling.
机译:介绍了用于膀胱监测的医疗植入式设备的设计和实现。加速度计的网络将植入外部膀胱肌肉和粘膜之间。还将压力传感器植入该粘膜下层以获得腔内压力读数。传感器系统由植入式电池供电,并提供用于数据和控制的无线通信链接。智能测量协议可使系统运行数周。传感器安装在可延长至250%的柔性印刷电路板上,以适应填充过程中膀胱器官组织的自然拉伸。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号