Assessment and comparison of radiation dose and image quality in multi-detector CT scanners in non-contrast head and neck examinations

机译:无头颈部检查中多探测器CT扫描仪的辐射剂量和图像质量的评估和比较

获取原文
           

摘要

Purpose To assess and compare radiation dose and image quality from non-contrast head and neck computed tomography (CT) examinations from four different multi-detector CT (MDCT) scanners. Material and methods Four CT scanners with different numbers of detector rows including one 4-MDCT, a 6-MDCT, a 16-MDCT, and a 64-MDCT were investigated. Common CT dose descriptors including volumetric CT dose index (CTDIv), dose length product (DLP), and the effective dose (ED), and image quality parameters include image noise, uniformity, and spatial resolution (SR) were estimated for each CT scanner with standard tools and methods. To have a precise comparison between CT scanners and related doses and image quality parameters, the ImPACT Q-factor was used. Results Minimum and maximum CTDIv, DLP, and ED in the head scan were 18 ± 3 and 49 ± 4 mGy, 242 ± 28 and 692 ± 173 mGy × cm, 0.46 ± 0.4 and 1.31 ± 0.33 mSv for 16-MDCT and 64-MDCT, respectively. And 16 ± 2 to 27 ± 3, 286 ± 127 to 645 ± 79 and 1.46 ± 0.65 to 3.29 ± 0.40 for neck scan, respectively. The Q-factor in head scan was 2.4, 3.3, 4.4 and 5.6 for 4-MDCT, 6-MDCT, 16-MDCT and 64-MDCT, respectively. The Q-factor in neck scan was 3.4, 4.6, 4.7 and 6.0 for 4-MDCT, 6-MDCT, 16-MDCT and 64-MDCT, respectively. Conclusions The results clearly indicate an increasing trend in the Q-factor from 4-MDCT to 64-MDCT units in both head and neck examinations. This increasing trend is due to a better SR and less noise of images taken and/or fewer doses in 64-MDCT.
机译:目的通过四个不同的多探测器CT(MDCT)扫描仪进行的非对比头颈计算机断层扫描(CT)检查,评估和比较辐射剂量和图像质量。材料和方法研究了四种具有不同检测器行数的CT扫描仪,包括一个4-MDCT,一个6-MDCT,一个16-MDCT和一个64-MDCT。为每个CT扫描仪估算了常见的CT剂量描述符,包括体积CT剂量指数(CTDIv),剂量长度乘积(DLP)和有效剂量(ED),以及图像质量参数,包括图像噪声,均匀性和空间分辨率(SR)使用标准工具和方法。为了在CT扫描仪和相关剂量以及图像质量参数之间进行精确比较,使用了ImPACT Q因子。结果头部扫描的CTDIv,DLP和ED的最小和最大分别为18±3和49±4 mGy,242±28和692±173 mGy×cm,16-MDCT和64-CT分别为0.46±0.4和1.31±0.33 mSv MDCT,分别。颈部扫描分别为16±2至27±3、286±127至645±79和1.46±0.65至3.29±0.40。 4-MDCT,6-MDCT,16-MDCT和64-MDCT的头部扫描Q值分别为2.4、3.3、4.4和5.6。 4-MDCT,6-MDCT,16-MDCT和64-MDCT的颈部扫描Q值分别为3.4、4.6、4.7和6.0。结论结果清楚地表明,在头部和颈部检查中,Q因子从4-MDCT单位增加到64-MDCT单位的趋势有所增加。这种增长趋势是由于在64-MDCT中具有更好的SR和所拍摄图像的噪点更少和/或剂量更少。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号