首页> 外文学位 >An Evaluation and Comparison of Beam Characteristics, Stray Radiation Room Surveys, Organ Dose, and Image Quality of Multiple Intra-Operative Imaging Devices for Orthopedic Lumbar Spinal Surgery.
【24h】

An Evaluation and Comparison of Beam Characteristics, Stray Radiation Room Surveys, Organ Dose, and Image Quality of Multiple Intra-Operative Imaging Devices for Orthopedic Lumbar Spinal Surgery.

机译:骨科腰椎手术的多种术中成像设备的束流特征,杂散辐射室测量,器官剂量和图像质量的评估和比较。

获取原文
获取原文并翻译 | 示例

摘要

Purpose: The overall purpose of this study was a comparison of radiation exposure for patients and staff during intra-operative imaging for orthopedic lumbar spine surgery. In order to achieve this, we: (1) Characterized each x-ray machine for physics performance, (2) Measured occupational radiation exposure inside the surgical suite for multiple intra-operative imaging devices utilizing currently in place clinical protocols for abdominal/spinal imaging, and (3) Measured specific organ doses for a phantom of three different Body Mass Indices (BMI) for each machine. We also compared the dose changes relative to changes in BMI as well as surgical image quality changes relative to BMI. This served as the majority of the first phase of a two phase project. The purpose of the second phase of the project will be to optimize scan parameters for surgical hardware placement in terms of image quality and organ dose for the devices that allow for modifications of scanner settings.;Materials and Methods: (1) X-Ray quality control meters were used to verify particular beam characteristics and additional information was calculated from the beam data. Both a small volume ionization chamber as well as Metal-Oxide-Semiconductor Field Effect Transistor (MOSFET) dosimeters were used to validate linear response of new design X-Ray tubes. (2) Both handheld ionization chamber survey meters as well as Geiger-Muller based personal dose meters were used to measure stray radiation for room surveys in locations representative of typical radiation worker positions during intra-operative imaging. (3) MOSFET dosimeters were placed in an adult male anthropomorphic phantom representing a normal BMI. 20 MOSFETs were used in nine organs with two small volume ion chambers used for skin surface dosimetry. Two additional layers of adipose equivalent material were progressively added to the phantom to represent BMI values of overweight and obese. Results: (1) The maximum tube potential, half value layer (HVL), effective energy, and soft tissue f-factor for each machine is as follows: IMRIS VISIUS iCT: 118.4 kVp, 7.66 mm Al, 53.64 keV, and 0.934 cGy/R; Mobis Airo: 122.3 kVp, 7.21 mm Al, 51.31 keV, and 0.925 cGy/R; Siemens ARCADIS Orbic 3D: 83 kVp, 7.12 mm Al, 32.76 keV, and 0.914 cGy/R; GE OEC 9900 Elite: 75 kVp, 4.25 mm Al, 46.6 keV, and 0.920 cGy/R. (2) The highest exposure rates measured during clinically implemented protocols for each scanner are as follows: IMRIS VISIUS iCT: 800 mR/hr; Mobis Airo: 6.47 R/hr; Siemens ARCADIS Orbic 3D: 26.4 mR/hr. (3) The effective dose per scan of each device for a full lumbar spine scan are as follows, for normal, overweight, and obese BMI, respectively: IMRIS VISIUS iCT: 12.00 +/- 0.30 mSv, 15.91 +/- 0.75 mSv, and 23.23 +/- 0.55 mSv; Mobius Airo: 5.90 +/- 0.25 mSv, 4.97 +/- 0.12 mSv, and 3.44 +/- 0.21 mSv; Siemens ARCADIS Orbic 3D: 0.30 +/- 0.03 mSv, 0.39 +/- 0.02 mSv, and 0.28 +/- 0.03 mSv; GE OEC 9900 Elite: 0.44 mSv, 0.77 mSv, and 1.14 mSv.;Conclusion: (1) The IMRIS VISIUS iCT i-Fluoro capable CT scanner and Mobius Airo mobile CT scanner have similar beam characteristics with significantly different tube parameter modulation protocols. Siemens ARCADIS Orbic 3D and GE OEC 9900 offer comparable beam characteristics but different imaging methods. All scanners performed within factory specifications. (2) The IMRIS VISIUS iCT should not be used in i-Fluoro mode for surgical procedures active during scanning due to the 1.42 cGy/s point dose rate in the beam field. The high exposure rate from the Mobius Airo is offset by short scan times and can be mitigated by ensuring enforcement of currently established radiation protection regulations and policies. Minimal stray radiation is measured from the Siemens ARCADIS Orbic 3D. (3) The differences in tube modulation of the CT scanners means the Mobius Airo offers a significantly reduced effective dose with increasing patient BMI over the IMRIS VISIUS iCT. Effective dose from the CT scanners varies as much as one to two orders of magnitude higher than the C Arms, but the Siemens ARCADIS Orbic 3D offers unusable image quality for patients with higher than normal BMI. Based off of physician reported usable surgical image quality of Mobius Airo, this device is recommended for continued integration and implementation during routine surgical procedures for patients of all BMI in orthopedic lumbar spine surgery.
机译:目的:本研究的总体目的是比较骨科腰椎手术患者术中成像期间患者和工作人员的放射线暴露。为了实现这一目标,我们:(1)对每台X射线机的物理性能进行表征,(2)使用当前就位的腹部/脊柱成像临床方案,为多个术中成像设备在手术室中测量到的职业辐射暴露,以及(3)为每台机器测量的三个不同人体质量指数(BMI)的模型的特定器官剂量。我们还比较了相对于BMI变化的剂量变化以及相对于BMI的手术图像质量变化。这是一个两阶段项目的第一阶段的大部分。该项目第二阶段的目的将是根据可修改扫描仪设置的设备的图像质量和器官剂量优化外科手术硬件放置的扫描参数。材料和方法:(1)X射线质量控制仪表用于验证特定的光束特性,并从光束数据中计算出其他信息。小型电离室和金属氧化物半导体场效应晶体管(MOSFET)剂量计均用于验证新型X射线管的线性响应。 (2)两种手持式电离室测量仪以及基于Geiger-Muller的个人剂量仪均用于测量杂散辐射,以便在代表术中成像过程中典型辐射工人位置的位置进行房间测量。 (3)将MOSFET剂量计放在代表正常BMI的成年男性拟人模型中。在九个器官中使用了20个MOSFET,并使用两个小体积离子室进行皮肤表面剂量测定。将两层额外的脂肪当量材料逐渐添加到体模中,以表示超重和肥胖的BMI值。结果:(1)每台机器的最大电子管电势,半值层(HVL),有效能量和软组织f因子如下:IMRIS VISIUS iCT:118.4 kVp,7.66 mm Al,53.64 keV和0.934 cGy / R; Mobis Airo:122.3 kVp,7.21毫米铝,51.31 keV和0.925 cGy / R;西门子ARCADIS Orbic 3D:83 kVp,7.12 mm Al,32.76 keV和0.914 cGy / R; GE OEC 9900 Elite:75 kVp,4.25 mm铝,46.6 keV和0.920 cGy / R。 (2)在临床实施方案中,对每台扫描仪测量的最高暴露率如下:IMRIS VISIUS iCT:800 mR / hr; Mobis Airo:6.47 R / hr;西门子ARCADIS Orbic 3D:26.4 mR / hr。 (3)对于正常,超重和肥胖的BMI,每台设备每次扫描的有效剂量分别如下:IMRIS VISIUS iCT:12.00 +/- 0.30 mSv,15.91 +/- 0.75 mSv,和23.23 +/- 0.55 mSv; Mobius Airo:5.90 +/- 0.25 mSv,4.97 +/- 0.12 mSv和3.44 +/- 0.21 mSv;西门子ARCADIS Orbic 3D:0.30 +/- 0.03 mSv,0.39 +/- 0.02 mSv和0.28 +/- 0.03 mSv; GE OEC 9900 Elite:0.44 mSv,0.77 mSv和1.14 mSv 。;结论:(1)具有IMRIS VISIUS iCT i-Fluoro功能的CT扫描仪和Mobius Airo移动CT扫描仪具有相似的光束特性,但管参数调制协议却大不相同。西门子ARCADIS Orbic 3D和GE OEC 9900具有可比的光束特性,但成像方法不同。所有扫描仪均在出厂规格范围内执行。 (2)由于光束场中的点剂量率为1.42 cGy / s,因此在i-Fluoro模式下不应将IMRIS VISIUS iCT用于扫描过程中有效的外科手术。短扫描时间抵消了Mobius Airo的高暴露率,并且可以通过确保执行当前建立的辐射防护法规和政策来缓解。最小杂散辐射是通过Siemens ARCADIS Orbic 3D测量的。 (3)CT扫描仪管调制的差异意味着,与IMRIS VISIUS iCT相比,Mobius Airo显着降低了有效剂量,并且患者的BMI有所增加。 CT扫描仪的有效剂量比C臂高出一到两个数量级,但是Siemens ARCADIS Orbic 3D为BMI高于正常水平的患者提供了无法使用的图像质量。基于医生报告的Mobius Airo可用的外科手术图像质量,推荐该设备在整形腰椎外科手术中所有BMI患者的常规手术过程中继续集成和实施。

著录项

  • 作者

    Womack, Kenneth Roland, II.;

  • 作者单位

    Duke University.;

  • 授予单位 Duke University.;
  • 学科 Medical imaging.
  • 学位 M.S.
  • 年度 2015
  • 页码 103 p.
  • 总页数 103
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

  • 入库时间 2022-08-17 11:52:51

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号