Power port contrast medium flushing and trapping: impact of temperature, an in vitro experimental study

机译:电源端口造影剂冲洗和捕集:温度的影响,一项体外实验研究

获取原文
           

摘要

Purpose: The use of totally implantable venous access devices (TIVADs) certified as "high pressure resistant" or "power port" has begun to spread worldwide as a safe procedure for power contrast injection. Owing to the thermo-rheological properties of the contrast media, the primary aim of this work is to present an in vitro experimental impact study concerning the impact of the temperature level on flushing efficiency after contrast medium injection. Moreover, we report experimental data that confirms the role of needle bevel orientation. The secondary aim is to answer the following questions: Is there significant device contrast medium trapping after contrast medium injection? Is saline flushing efficient? And, finally, is it safe to inject contrast medium through an indwelled port catheter?Results: The experimental results show that in addition to hydrodynamics, temperature is a key parameter for the efficiency of device flushing after contrast medium injection. It appears that this is the case when the cavity is incompletely rinsed after three calibrated flushing volumes of 10 mL saline solution, even by using the Huber needle bevel opposite to the port exit. This leads to a potentially important trapped volume of contrast medium in the port, and consequently to the possibility of subsequent salt precipitates and long term trisubstituted benzene nuclei delivery that might impair the solute properties, which may be further injected via the power port later on.Conclusion: We thus suggest, in TIVADS patients, the use of a temporary supplementary intravenous line rather than the port to perform contrast medium injections in daily radiology routine practice.
机译:目的:认证为“耐高压”或“电源端口”的完全植入式静脉通路装置(TIVAD)的使用已开始普及,作为安全的动力对比注射方法。由于造影剂的热流变特性,这项工作的主要目的是针对注射造影剂后温度水平对冲洗效率的影响进行体外实验研究。此外,我们报告了实验数据,证实了针斜角定位的作用。第二个目的是回答以下问题:注射造影剂后是否存在大量的器械造影剂滞留?盐水冲洗有效吗?最后,通过留置口导管注入造影剂是否安全?结果:实验结果表明,除了流体力学以外,温度是注射造影剂后冲洗设备效率的关键参数。似乎是这种情况,即使在使用三个校准的冲洗体积的10 mL盐溶液后,即使使用与端口出口相对的Huber针斜面,也无法完全冲洗腔体。这导致造影剂在端口中的潜在重要捕集体积,并因此导致随后的盐沉淀和长期的三取代苯核传递可能损害溶质的可能性,随后可能会通过电源端口将其注入。结论:因此,我们建议在日常放射学常规操作中,在TIVADS患者中,使用临时的补充静脉管线而不是端口进行造影剂注射。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号