首页> 外文期刊>International Journal of Medical Physics, Clinical Engineering and Radiation Oncology >Dosimetric Comparison of Craniospinal Axis Irradiation (CSI) Treatments Using Helical Tomotherapy, SmartarcTM, and 3D Conventional Radiation Therapy
【24h】

Dosimetric Comparison of Craniospinal Axis Irradiation (CSI) Treatments Using Helical Tomotherapy, SmartarcTM, and 3D Conventional Radiation Therapy

机译:使用螺旋断层扫描,SmartarcTM和3D常规放射疗法对颅骨脊髓轴照射(CSI)治疗进行剂量学比较

获取原文
           

摘要

Purpose: Craniospinal axis irradiation (CSI) is a method of treating various central nervous system malignancies. The large target volume typically includes entire spinal cord and whole brain. Dosimetric comparison was performed between tomotherapy, volumetric modulated arc therapy (VMAT), and 3D conformal radiation therapy (3D-CRT) for CSI. Methods and Materials: Five (n = 5) CSI patients were planned using 3D-CRT, VMAT, and tomotherapy (normalized such that 95% of PTV received at least 23.4 Gy in 13 fractions). Plans were compared using PTV conformity number (CN) and homogeneity index (HI), normal tissue (NT) dose statistics, integral dose, and treatment time. Results: On average, tomotherapy plans showed higher CN (0.932 vs. 0.860 and 0.672 for SmartArc and 3D-CRT). In terms of HI, VMAT plans consistently showed better dose homogeneity (1.07 vs. 1.15 and 1.13 for tomotherapy and 3D-CRT). SmartArc delivered lower maximum dose for majority of NT, but higher mean dose. 3D-CRT plans delivered higher maximum dose but lower mean dose to NT. Conclusions: SmartArc treatments achieved better PTV homogeneity and reduced maximum dose to NT. Tomotherapy showed better target conformity, but 3D-CRT was shown to reduce mean dose to NT. Integral doses were similar between treatment modalities, but tomotherapy treatment times were much longer.
机译:目的:颅颈椎轴照射(CSI)是一种治疗各种中枢神经系统恶性肿瘤的方法。大目标体积通常包括整个脊髓和整个大脑。在CSI的tomotherapy,体积调制弧光疗法(VMAT)和3D保形放射疗法(3D-CRT)之间进行了剂量学比较。方法和材料:计划使用3D-CRT,VMAT和tomotherapy(正常化以使95%的PTV在13个分数中至少获得23.4 Gy)计划治疗5例(n = 5)CSI患者。使用PTV合格数(CN)和均匀性指数(HI),正常组织(NT)剂量统计数据,积分剂量和治疗时间对计划进行比较。结果:平均而言,断层治疗计划显示较高的CN(SmartArc和3D-CRT的0.932对0.860和0.672)。就HI而言,VMAT计划始终显示出更好的剂量均匀性(X线断层扫描和3D-CRT的剂量均匀性为1.07对1.15和1.13)。 SmartArc对大部分NT的最大剂量较低,但平均剂量较高。 3D-CRT计划向NT提供了更高的最大剂量,但降低了平均剂量。结论:SmartArc治疗获得了更好的PTV同质性并减少了NT的最大剂量。断层扫描显示更好的靶标顺应性,但3D-CRT可以降低NT的平均剂量。在两种治疗方式之间,整体剂量相似,但是tomotherapy的治疗时间更长。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号