Application of CZE Method in Routine Analysis for Determination of B-Complex Vitamins in Pharmaceutical and Veterinary Preparations

机译:CZE方法在常规分析中测定兽药制剂中B类维生素的含量

获取原文
           

摘要

A competitive CZE method for quality control analysis of multivitamin preparations and veterinary products containing B-group vitamins was developed. Vitamins of interest are thiamine hydrochloride (B1), thiamine monophosphate chloride (B1a), riboflavine (B2), riboflavine-5′monophosphate (B2a), nicotinamide (B3), d-pantothenic acid calcium salt (B5), pyridoxine hydrochloride (B6), folic acid (B9), and 4-aminobenzoic acid (B10). These analytes were separated optimizing the experimental conditions in 20 mM tetraborate bufferpH=9.2as a BGE (background electrolyte), on a Beckman P/ACE System MDQ instrument, using uncoated fused silica capillary. The effective capillary length was of 49.5 cm,I.D.=50 μm, the applied voltage 20 kV and the temperature 25∘C. Detection was performed by a diode array detector at 214 nm for all vitamins except B5(190 nm) and B2a(260 nm). Separation time was about 9 min. After experimental conditions optimization, the proposed method was validated. Precision of migration time and corrected peak area, linearity range, LOD and LOQ, accuracy (recovery), robustness, and ruggedness were evaluated for each analyte demonstrating the good reliability of the method. Analyses of the pharmaceutical real samples were performed and confirmed the versatility of this method.
机译:开发了一种竞争性的CZE方法,用于分析多种维生素制剂和含有B组维生素的兽药的质量。感兴趣的维生素为盐酸硫胺素(B1),氯化硫胺素一磷酸盐(B1a),核黄素(B2),核黄素5'一磷酸(B2a),烟酰胺(B3),d-泛酸钙盐(B5),吡ido醇盐酸盐(B6) ),叶酸(B9)和4-氨基苯甲酸(B10)。在Beckman P / ACE System MDQ仪器上,使用未涂覆的熔融石英毛细管,在作为BGE(背景电解质)的20 mM四硼酸盐缓冲液pH = 9.2中优化实验条件,分离了这些分析物。有效毛细管长度为49.5 cm,内径= 50μm,施加电压20 kV,温度25 andC。用二极管阵列检测器在214 nm处检测除B5(190 nm)和B2a(260 nm)以外的所有维生素。分离时间约为9分钟。经过实验条件的优化,对提出的方法进行了验证。评估了每种分析物的迁移时间和校正的峰面积,线性范围,LOD和LOQ的准确性,准确性(回收率),稳健性和耐用性,证明了该方法的良好可靠性。进行了药物真实样品的分析,并证实了该方法的多功能性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号