T-wave oversensing can cause inappropriate implantable cardioverter-defibrillator (ICD) therapies that are difficult to correct. Remote monitoring allows follow-up of ICD patients without visiting the hospital and can help in early detection of any malfunctions. We describe the case of a patient who experienced inappropriate antitachycardia pacing therapy due to T-wave oversensing; the problem was promptly detected by remote monitoring and corrected by device reprogramming.
机译:T波感应过度会导致难以纠正的不合适的植入式心脏除颤器(ICD)治疗。远程监控可以对ICD患者进行随访,而无需去医院就诊,并且可以帮助及早发现任何故障。我们描述了由于T波过度感应而经历了不合适的抗心动过速起搏治疗的患者。该问题可通过远程监控迅速发现,并通过设备重新编程予以纠正。