首页> 美国卫生研究院文献>Indian Pacing and Electrophysiology Journal >T-wave Oversensing with Inappropriate Therapy in Remote Monitoring
【2h】

T-wave Oversensing with Inappropriate Therapy in Remote Monitoring

机译:远程监测中治疗不当的T波感应过度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

T-wave oversensing can cause inappropriate implantable cardioverter-defibrillator (ICD) therapies that are difficult to correct. Remote monitoring allows follow-up of ICD patients without visiting the hospital and can help in early detection of any malfunctions. We describe the case of a patient who experienced inappropriate antitachycardia pacing therapy due to T-wave oversensing; the problem was promptly detected by remote monitoring and corrected by device reprogramming.
机译:T波感应过度会导致难以纠正的不合适的植入式心脏除颤器(ICD)治疗。远程监控可以对ICD患者进行随访,而无需去医院就诊,并且可以帮助及早发现任何故障。我们描述了由于T波过度感应而经历了不适当的抗心动过速起搏治疗的患者。该问题可通过远程监控迅速发现,并通过设备重新编程予以纠正。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号