首页> 美国卫生研究院文献>Journal of Arrhythmia >A case of inappropriate implantable cardioverter defibrillator therapy induced by T-wave oversensing due to hyperkalemia
【2h】

A case of inappropriate implantable cardioverter defibrillator therapy induced by T-wave oversensing due to hyperkalemia

机译:高钾血症引起的T波过度感应导致植入式心脏复律除颤器治疗不当的情况

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

There have been reports of hyperkalemia-induced T-wave oversensing in patients with implantable cardioverter defibrillators (ICDs). However, a comparison of T-wave amplitudes and morphologies between the surface 12-lead electrocardiogram (ECG) and ICD electrogram has not been reported. We present the case of a 70-year-old man who received inappropriate ICD shocks due to hyperkalemia-induced T-wave oversensing. The T-wave amplitudes on both the ICD electrogram and 12-lead ECG corresponded and normalized after normalization of the potassium level.
机译:有报道称高钾血症引起的植入式心脏复律除颤器(ICD)患者的T波过度感应。但是,尚未报道在表面12导联心电图(ECG)和ICD电描记图之间比较T波幅度和形态。我们介绍了一个案例,该例是一名70岁的男子,由于高钾血症引起的T波过度感应而收到了不适当的ICD电击。钾水平归一化后,ICD电泳图和12导联心电图上的T波振幅都对应并归一化。