首页> 外文期刊>Annual Research & Review in Biology >Reaction Force Measurement in Milwaukee Brace during Daily Activities-Introduction of a New Measurement System
【24h】

Reaction Force Measurement in Milwaukee Brace during Daily Activities-Introduction of a New Measurement System

机译:日常活动中密尔沃基支撑中的反作用力测量-一种新的测量系统的介绍

获取原文
       

摘要

Aims: The reaction forces of the Milwaukee Brace’s (MB’s) several areas were measured during normal daily activities (24 hr) by a new measurement device improved in order to force acquisition by groups of sensors distributed in all of the MB’s internal areas. Methodology: In the current study, a new measurement system using Force Sensitive Resistance (FSRs) distributed on the different areas of MB for scoliosis patients. Sensitivity analysis has been fulfilled to compare two different sensors arrangements placed in the basket of MB. A reliability analysis has been carried out for normal activities including sitting, walking and sleeping. Two healthy volunteers and two scoliosis patients were nominated to force acquisition from their MB during 23 hours of daily activities. Results: In the sensitivity analysis, maximum value of Cronbach’s Alpha was 0.775 for sitting at basket lateral side. In the reliability analysis, the maximum and minimum difference between the force values of the two arrangements is 1.37(N) and 0.19(N) respectively related to the sitting test. The tests for each subject commenced with 9-11 hours of sleeping and continued for 12-14 hours in which main activities were sitting and walking. The forces average values of each part of the tests were reported for different areas of MBs. The mean, median, mode, standard deviation and ratio of standard deviation on the mean have been calculated for each brace region for both time durations. Conclusion: The brace different areas forces for sleeping and day time activities had normal distribution. The force value of the brace regions having direct contact with mattress surface was increased during sleeping with respect to day time activities. Other regions forces were decreased or remained constant in this time duration. The forces detected from the lumbar pad during most of the test time were constant.
机译:目的:通过改进的新型测量设备,在正常的日常活动(24小时)期间,对密尔沃基支撑(MB)几个区域的反作用力进行了测量,以迫使分布在MB内部所有区域的传感器组进行采集。方法:在当前的研究中,针对脊柱侧弯患者,使用力敏感电阻(FSR)分布在MB的不同区域的新测量系统。已完成灵敏度分析,以比较放置在MB篮中的两个不同的传感器布置。已针对包括坐,走和睡觉在内的正常活动进行了可靠性分析。提名两名健康志愿者和两名脊柱侧弯患者在每天23小时的活动中强迫他们的MB进行采集。结果:在敏感性分析中,坐在篮筐外侧的克朗巴赫Alpha最大值为0.775。在可靠性分析中,两种布置的力值之间的最大和最小差分别为1.37(N)和0.19(N)。每个受试者的测试从9-11个小时的睡眠开始,持续12-14个小时,主要活动是坐着和走路。报告了MB的不同区域的测试各部分的力平均值。对于两个持续时间的每个支撑区域,均已计算出平均值,中位数,众数,标准偏差和标准偏差与平均值的比率。结论:支撑不同区域的睡眠力和白天活动力呈正态分布。就白天活动而言,在睡眠期间增加了与床垫表面直接接触的支撑区域的力值。在此时间段内,其他区域的力减小或保持恒定。在大多数测试时间内,从腰垫检测到的力是恒定的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号