首页> 外文期刊>BMC Urology >Evaluation of a rapid quantitative determination method of PSA concentration with gold immunochromatographic strips
【24h】

Evaluation of a rapid quantitative determination method of PSA concentration with gold immunochromatographic strips

机译:金免疫层析条快速定量测定PSA浓度的方法评价

获取原文
           

摘要

Prostate cancer remains the most common cancer in men. Qualitative or semi-quantitative immunochromatographic measurements of prostate specific antigen (PSA) have been shown to be simple, noninvasive and feasible. The aim of this study was to evaluate an optimized gold immunochromatographic strip device for the detection of PSA, in which the results can be analysed using a Chromogenic Rapid Test Reader to quantitatively assess the test results. This reader measures the reflectance of the signal line via a charge-coupled device camera. For quantitative analysis, PSA concentration was computed via a calibration equation. Capillary blood samples from 305 men were evaluated, and two independent observers interpreted the test results after 12?min. Blood samples were also collected and tested with a conventional quantitative assay. Sensitivity, specificity, positive and negative predictive values, and accuracy of the PSA rapid quantitative test system were 100, 96.6, 89.5, 100, and 97.4?%, respectively. Reproducibility of the test was 99.2, and interobserver variation was 8?% with a false positive rate of 3.4?%. The correlation coefficient between the ordinary quantitative assay and the rapid quantitative test was 0.960. The PSA rapid quantitative test system provided results quickly and was easy to use, so that tests using this system can be easily performed at outpatient clinics or elsewhere. This system may also be useful for initial cancer screening and for point-of-care testing, because results can be obtained within 12?min and at a cost lower than that of conventional quantitative assays.
机译:前列腺癌仍然是男性中最常见的癌症。前列腺特异性抗原(PSA)的定性或半定量免疫色谱测定已被证明是简单,无创且可行的。这项研究的目的是评估一种用于PSA检测的优化的金免疫层析带装置,其中可以使用生色快速检测仪对结果进行分析,以定量评估检测结果。该阅读器通过一个电荷耦合设备的摄像头测量信号线的反射率。对于定量分析,通过校准方程计算PSA浓度。对305名男性的毛细血管血样进行了评估,两名独立的观察员在12分钟后对测试结果进行了解释。还收集血样并通过常规定量测定法进行测试。 PSA快速定量测试系统的灵敏度,特异性,阳性和阴性预测值和准确性分别为100%,96.6%,89.5%,100%和97.4%。测试的可重复性为99.2,观察者之间的差异为8%,假阳性率为3.4%。普通定量分析与快速定量检测的相关系数为0.960。 PSA快速定量测试系统可以快速提供结果,并且易于使用,因此可以在门诊诊所或其他地方轻松进行使用该系统的测试。该系统对于最初的癌症筛查和即时检验也可能有用,因为可以在12分钟内获得结果,而且成本要低于传统的定量分析方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号