As part of the programme to develop low-cost instruments for use in educational institutions, we have developed a heat-flux differential scanning calorimeter (HF-DSC). We have used low cost components for building the parts of the calorimeter and they were fabricated by the local industries. The paper describes the design and construction of the calorimeter and measurement electronics in detail. The calorimeter was calibrated for temperature and enthalpy measurements using high purity metal standards. The results indicate that the on-set temperatures of thermal events can be measured with an accuracy of ± 0.7 °C and the enthalpy values within ± 3%. The sensitivity of the instrument is about 4 to 5 times lower than that obtained in a typical commercial DSC unit.
机译:作为开发用于教育机构的低成本工具的计划的一部分,我们开发了一种热量差动扫描量热计(HF-DSC)。 我们使用了低成本组件来构建量热计的部件,并由当地行业制造。 本文详细介绍了量热计和测量电子的设计和构造。 使用高纯度金属标准校准量热计进行温度和焓测量。 结果表明,热事件的内置温度可以以±0.7°C的精度测量,±3%内的焓值。 仪器的灵敏度比在典型的商业DSC单元中获得的约4至5倍。