Influence of Socket-shield technique on the biomechanical response of dental implant: three-dimensional finite element analysis

机译:插座屏蔽技术对牙科植入物生物力学响应的影响:三维有限元分析

获取原文
获取原文并翻译 | 示例
           

摘要

This study evaluated the bone microstrain, displacement, and stress distribution according to the surgical technique (conventional or Socket-shield) and evaluation period (immediately after implant installation or after healing). Each condition was modeled for the finite element analysis, totaling four groups, with a morse-taper implant and a cemented prosthesis. The maximum displacement, von Mises stress, and bone microstrain yielded higher values during the immediate stage, without a difference between Socket-shield and conventional treatments. The use of the Socket-shield technique does not negatively impact the biomechanical behavior of an implant-supported prosthesis immediately after healing from the implant installation.