Pneumatic muscle actuators (PMAs) offer significant advantages over more traditional actuators, which make them prime candidates in rehabilitation devices. A dynamic test station (DTS) is modified to demonstrate the use of a PMA for this application. The DTS includes two dynamic systems: a PMA and a DC servomotor. An overall transfer function was developed utilising characterisation data for the PMA and DC servomotor. A Tustin (bilinear) transform was performed on the overall transfer function to obtain a discrete time system. Model parameters were optimised and used to generate input voltage profiles that achieve isokinetic (constant velocity) task specifications. Percent root mean square error values (PRMSE) between the actual and ideal profiles were used to evaluate the accuracy of this method in achieving isokinetic displacement. For PMA pressures (in kPa) of 150, 350 and 550 PRMSE were 7.80, 5.40 and 2.76, respectively.
机译:气动肌肉致动器(PMA)与更传统的致动器相比具有明显的优势,这使其成为康复设备的主要候选产品。修改了动态测试站(DTS),以演示此应用程序对PMA的使用。 DTS包括两个动态系统:PMA和DC伺服电动机。利用PMA和DC伺服电机的特性数据开发了整体传递函数。对整个传递函数执行Tustin(双线性)变换,以获得离散时间系统。优化了模型参数,并将其用于生成实现等速(恒速)任务规范的输入电压曲线。实际和理想曲线之间的均方根误差百分比(PRMSE)用于评估此方法在实现等速位移时的准确性。对于150、350和550的PMA压力(以kPa为单位),PRMSE分别为7.80、5.40和2.76。