Variation of foot-to-foot resistance measured by body fat analysers with electrode geometry and contact area

机译:人体脂肪分析仪测量的脚对脚电阻随电极几何形状和接触面积的变化

获取原文
获取原文并翻译 | 示例

摘要

This work showed that the resistance measurement can be influenced by characteristics of the measuring circuit and by the size of current electrodes, while the size of voltage electrodes has less effect. An increase in frequency from 40 to 50 kHz reducing the resistance by about 10 fl is also observed. The relative position of feet with metal electrodes has a major effect on the reproducibility of resistance measurement; when feet were moved backward by 5 cm, the foot-to-foot resistance increased by 20-30 Ω, due to a rise in length of current path under the feet.
机译:这项工作表明,电阻测量会受到测量电路特性和电流电极尺寸的影响,而电压电极的尺寸影响较小。还观察到频率从40 kHz增加到50 kHz,电阻降低了约10 fl。支脚与金属电极的相对位置对电阻测量的可重复性有重要影响。当脚向后移动5 cm时,由于脚下电流路径的长度增加,脚对脚的电阻增加了20-30Ω。