首页> 外文期刊>IEEE Transactions on Biomedical Engineering >A moving-actuator type electromechanical total artificial heart. I. Linear type and mock circulation experiments
【24h】

A moving-actuator type electromechanical total artificial heart. I. Linear type and mock circulation experiments

机译:移动致动器型机电全人工心脏。 I.线性类型和模拟循环实验

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A novel type of motor-driven total artificial heart system with a moving-actuator mechanism has been developed. The prototype system consists of a brushless DC motor inside a rolling cylinder, two arcshaped pusher plates, and two polyurethane sacs. The moving-actuator-type electromechanical pump has the structural advantages of small size and light weight, as compared to other motor-driven pumps with fixed actuator mechanisms. Results of mock circulation tests with a cardiac output of 9 L/min at an aortic pressure of 120 mm Hg and a heart rate of 120 b/m are reported.
机译:已经开发了一种新型的带有运动致动器机构的电机驱动的全人工心脏系统。原型系统由位于滚动缸内的无刷直流电动机,两个弧形推板和两个聚氨酯囊组成。与其他具有固定致动器机构的电动泵相比,该移动致动器型机电泵具有体积小,重量轻的结构优势。据报道,在主动脉压力为120 mm Hg且心律为120 b / m时,心输出量为9 L / min的模拟循环测试结果。