Bi-Modal Arterial Compliance Probe for Calibration-Free Cuffless Blood Pressure Estimation

机译:双模态动脉顺应性探头,用于免校正无袖带血压估算

获取原文
获取原文并翻译 | 示例
           

摘要

Objective: We propose a calibration-free method and system for cuffless blood pressure (BP) measurement from superficial arteries. A prototype device with bi-modal probe arrangement was designed and developed to estimate carotid BP - an indicator of central aortic pressure. Methods: Mathematical models relating BP parameters of an arterial segment to its dimensions and local pulse wave velocity (PWV) are introduced. A bi-modal probe utilizing ultrasound and photoplethysmograph sensors was developed and used to measure diameter values and local PWV from the carotid artery. Carotid BP was estimated using the measured physiological parameters without any subject- or population-specific calibration procedures. The proposed cuffless BP estimation method and system were tested for accuracy, usability, and for potential utility in hypertension screening, on a total of 83 subjects. Results: The prototype device demonstrated its capability of detecting beat-by-beat arterial dimensions and local PWV simultaneously. Carotid diastolic BP (DBP) and systolic BP (SBP) were estimated over multiple cardiac cycles in real-time. The absolute error in carotid DBP was <;10 mmHg in 82% cases, and root-mean-square-error = 8.3 mmHg. Consistent with the theory, estimated SBP at the carotid site was lower than the reference brachial SBP. ROC curves obtained for hypertension screening analysis revealed an area under the curve ≥0.8 for both carotid SBP and DBP values, illustrating the potential for using the developed method in hypertension screening. Conclusion: The feasibility of calibration-free, cuffless BP measurement at an arterial site of interest was demonstrated with a level of acceptable accuracy. The study also demonstrated the potential utility of the proposed method and system in hypertension screening and local evaluation of arterial stiffness indices. Significance: Novel approach for calibration-free cuffless BP estimation; a potential tool for local BP measurement and hypertension screening.
机译:目的:我们提出了一种用于浅表动脉无袖带血压(BP)测量的免校准方法和系统。设计并开发了一种具有双峰探头布置的原型设备,以估计颈动脉BP(颈动脉BP),这是中心主动脉压力的指标。方法:介绍了将动脉节段的BP参数与其尺寸和局部脉搏波速度(PWV)相关联的数学模型。开发了一种利用超声波和光电容积描记器传感器的双峰探头,并将其用于测量颈动脉的直径值和局部PWV。在没有任何受试者或人群特异性校准程序的情况下,使用测量的生理参数估算颈动脉血压。在总共83名受试者中,对提出的无袖带BP估计方法和系统进行了准确性,可用性和在高血压筛查中的潜在效用的测试。结果:该原型设备展示了其能够同时检测逐搏动脉尺寸和局部PWV的能力。实时评估多个心动周期的颈动脉舒张压(DBP)和收缩压(SBP)。在82%的病例中,颈动脉DBP的绝对误差<; 10 mmHg,并且均方根误差= 8.3 mmHg。与理论相符,颈动脉部位的SBP估计值低于参考臂SBP。用于高血压筛查分析的ROC曲线显示,对于颈动脉SBP和DBP值,曲线下的面积均≥0.8,这说明了将开发的方法用于高血压筛查的潜力。结论:在目标动脉部位进行无标定,无袖带BP测量的可行性已被证明具有可接受的准确性。该研究还证明了所提出的方法和系统在高血压筛查和动脉僵硬度指数的局部评估中的潜在实用性。启示:一种无需校准的无袖带BP估计新方法;一个潜在的局部BP测量和高血压筛查工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号