We have developed superconducting single-photon imaging X-ray detectors with an energy resolution of 54 eV at 6 keV and a spatial resolution of 1 μm over an effective length of 40 μm. They utilize a current-sensitive low-noise preamplifier with a dc voltage bias. It has a signal bandwidth of 300 kHz, current noise of in=0.26 pA/√(Hz) and voltage noise of en=0.5 nV/√(Hz) with an input capacitance of 200 pF under operating conditions. Injected pulses with a charge Q=3.7·106 electrons have been measured with a standard deviation σQ=3400 electrons, corresponding to an electronic noise of 13 eV at 6 keV
机译:我们已经开发了在6 keV时的能量分辨率为54 eV,在有效长度为40μm的情况下空间分辨率为1μm的超导单光子成像X射线探测器。它们利用具有直流电压偏置的电流敏感型低噪声前置放大器。在工作条件下,输入电容为200 pF,信号带宽为300 kHz,电流噪声为in = 0.26 pA /√(Hz),电压噪声为en = 0.5 nV /√(Hz)。已测量了电荷为Q = 3.7·106电子的注入脉冲,其标准偏差为σQ= 3400电子,对应于6 keV时13 eV的电子噪声