aSi-EPID transit signal calibration for dynamic beams: a needful step for the IMRT in vivo dosimetry

机译:动态光束的aSi-EPID传输信号校准:IMRT体内剂量测定的必要步骤

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This work reports a method based on correlation functions to convert EPID transit signals into in vivo dose values at the isocenter point, Diso, of dynamic IMRT beams supplied by Varian linac. Dose reconstruction for intensity-modulated beams required significant corrections of EPID response, due to the X-ray component transmitted through multileaf collimator. The algorithm was formulated using a set of simulated IMRT beams. The beams were parameterized by means of a fluence inhomogeneity index, FI, introduced to describe the degree of beam modulation with respect to open beams. This way, all dosimetric parameters involved in Diso reconstruction algorithm, such as the correlation functions, the correction factor for EPID to phantom distance and the modulated tissue maximum ratios, were determined as a function of the FI index. Clinical IMRT beams were used to irradiate a homogeneous phantom, and for each beam, the agreement between the reconstructed dose, Diso, and the dose computed by TPS, Diso,TPS, was well within 5 %. Moreover, the average ratios, R, between the Diso, and Diso,TPS, resulted equal to 1.002 ± 0.030. Thirty-five IMRT fields of 5 different patients undergoing radiotherapy for head–neck tumors were tested and the results were displayed on a computer screen after 2 min from the end of the treatment. However, 350 in vivo tests supplied an average ratio R equal to 1.004 ± 0.040. The in vivo dosimetry procedure here presented is among the objectives of a National Project financially supported by the Istituto Nazionale di Fisica Nucleare for the development of in vivo dosimetry procedures (Piermattei et al. in Nucl Instrum Methods Phys Res B 274:42–50, ) connected to the Record–Verify system of the radiotherapy center.
机译:这项工作报告了一种基于相关函数的方法,该方法将EPID转换信号转换为Varian直线加速器提供的动态IMRT光束的等中心点Diso处的体内剂量值。由于通过多叶准直仪传输的X射线分量,用于强度调制光束的剂量重建需要EPID响应的显着校正。该算法是使用一组模拟IMRT光束制定的。光束通过注量不均匀性指数FI进行参数化,该指数被引入以描述相对于开放光束的光束调制度。通过这种方式,确定了Diso重建算法中涉及的所有剂量参数,例如相关函数,EPID到体模距离的校正因子以及调制的组织最大比率,都取决于FI指数。临床IMRT光束用于照射均匀的体模,对于每个光束,重建剂量Diso与TPS计算的剂量Diso,TPS之间的一致性都在5%之内。而且,Diso和Diso,TPS之间的平均比率R等于1.002±0.030。对5例接受放射治疗的头颈部肿瘤患者的35个IMRT场进行了测试,并将结果从治疗结束2分钟后显示在计算机屏幕上。但是,350个体内测试提供的平均比率R等于1.004±0.040。此处介绍的体内剂量测定方法是国家科学基金会资助的国家项目的目标之一,该计划的目的是开发体内剂量测定方法(Piermattei等人,在Nucl Instrum Methods Phys Res B 274:42–50中, )连接到放疗中心的Record-Verify系统。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号