首页> 美国卫生研究院文献>The Journal of Automatic Chemistry >Validation of an HPLC Method for Determination of Bisphenol-A Migration from Baby Feeding Bottles
【2h】

Validation of an HPLC Method for Determination of Bisphenol-A Migration from Baby Feeding Bottles

机译:HPLC法测定双酚A从婴儿奶瓶中迁移的验证

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A simple and economic high-performance liquid chromatography (HPLC-UV-Vis) analytical method was validated for the quantitation of specific Bisphenol-A migration from baby feeding bottles. Overall and specific migration assays were done with different food simulating matrices using the filling method. Good linearity was obtained over the concentration range of 0.01–0.6 mg/kg. The limit of detection (LOD) and limit of quantification (LOQ) were 0.004 and 0.010 mg/kg, respectively. The repeatability of the method (%RSD, n=10) was between 89.5 and 99.0%, while recovery ranged from 83.2 to 98.4%. The method was applied to specific migration assays from baby feeding bottles purchased from different plastic producers in Colombia. The results show that, in a first migration assay, Bisphenol-A was not detectable in all samples. In a second migration test, Bisphenol-A concentrations were higher than the most restricted limit (0.05 mg/kg) with ethanol 95% and isooctane as food simulants.
机译:验证了一种简单经济的高效液相色谱(HPLC-UV-Vis)分析方法,用于定量从婴儿奶瓶中迁移出的特定双酚A。使用填充方法,对不同的食品模拟基质进行了总体迁移和特异性迁移分析。在0.01-0.6μmg/ kg的浓度范围内获得了良好的线性。检测限(LOD)和定量限(LOQ)分别为0.004和0.010μmg/ kg。该方法的重复性(%RSD,n = 10)在89.5%至99.0%之间,而回收率在83.2%至98.4%之间。该方法用于从哥伦比亚不同塑料生产商处购买的婴儿奶瓶的特定迁移分析。结果表明,在首次迁移分析中,并非在所有样品中都检测到双酚A。在第二次迁移测试中,使用95%的乙醇和异辛烷作为食品模拟物,双酚A的浓度高于最高限制限值(0.05μmg/ kg)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号