Patient’s Radiation Exposure in Coronary Angiography and Angioplasty: The Impact of Different Projections

机译:冠状动脉造影和血管成形术中患者的放射线暴露:不同投影的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Introduction: We aimed to determine angiography projections with lower Dose Area Product (DAP) rate by measuring the mean DAP and fluoroscopy times in coronary angiography (CAG) and percutaneous coronary intervention (PCI) and calculating DAP rate in different projections. >Methods: DAP and fluoroscopy times were measured in all employed projections in real-time in 75 patients who underwent CAG or PCI by a single cardiologist in Madani Cardiovascular University Hospital (45 in CAG group and 30 in PCI group). DAP rate was calculated in both groups and in all projections. The projections with highest and lowest DAP rate were determined. >Results: Mean DAP was 436.73±315.85 dGy×cm2 in CAG group and 643.26±359.58 dGy×cm2 in PCI group. The projection 40° LAO/0° had the highest DAP rate in CAG group (28.98 dGy×cm2/ sec) and it was highest in 20° RAO/30° CR in PCI group (29.83 dGy×cm2/sec). The latter projection was also the most employed projection in PCI group. >Conclusion: The amount of radiation dose in this study is in consistent with the previous reports. Specific angiographic projections expose patients to significantly higher radiation and they should be avoided and replaced by less irradiating projections whenever possible.
机译:>简介:我们旨在通过测量冠状动脉造影(CAG)和经皮冠状动脉介入治疗(PCI)中的平均DAP和透视时间,并计算DAP率来确定具有较低剂量面积积(DAP)率的血管造影预测。不同的预测。 >方法:由Madani心血管大学医院的一名心脏病专家对75位行CAG或PCI的患者进行实时测量,以测量所有采用的投影法中的DAP和透视时间(CAG组为45名,PCI组为30名) )。计算两组和所有预测的DAP率。确定了DAP率最高和最低的预测。 >结果:CAG组平均DAP为436.73±315.85 dGy×cm 2 ,PCI组平均DAP为643.26±359.58 dGy×cm 2 。投影40°LAO / 0°在CAG组中DAP率最高(28.98 dGy×cm 2 / sec),在PCI组20°RAO / 30°CR中最高(29.83 dGy ×cm 2 / sec)。后者的预测也是PCI组中使用最多的预测。 >结论:本研究中的辐射剂量与以前的报告一致。特定的血管造影投影会使患者暴露于明显更高的辐射下,应避免使用它们,并在可能的情况下用较少辐射的投影代替。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号