Effects of ball head diameter and stem neck shape in range of motion after total hip arthroplasty: A simulation study

机译:球头直径和茎颈形状在总髋关节置换术后运动范围内的影响:模拟研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Simulations using three-dimensional computer-aided design were performed with the various conditions to research the movement distance of the ball head and total range of motion until impingement and dislocation (ROM(t)) which involved in joint stability. Although a large ball head increases the entire ROM(t) and oscillation angle, it decreases stability transitioning from impingement to dislocation. Conversely, though large neck diameter increases the simulated stability from impingement to dislocation, it reduces the ROM(t) and oscillation angle. ROM(t) decreases with change in neck shape and is greatly affected by the presence of osteophytes, even if the head–neck ratio remains the same.
机译:使用三维计算机辅助设计进行使用各种条件来研究球头的移动距离和涉及关节稳定性的冲击和错位(ROM(T))。尽管大的球头增加了整个ROM(T)和振荡角度,但它会降低从冲击到位错的稳定性转换。相反,虽然大颈部直径从冲击到位脱位增加模拟稳定性,但它减少了ROM(T)和振荡角度。 ROM(t)随颈部形状的变化而减小,并且即使头颈比保持不变,也会受到骨赘的存在。