首页> 美国卫生研究院文献>other >Mapping antiretroviral drugs in tissue by IR-MALDESI MSI coupled to the Q Exactive and comparison with LC-MS/MS SRM Assay
【2h】

Mapping antiretroviral drugs in tissue by IR-MALDESI MSI coupled to the Q Exactive and comparison with LC-MS/MS SRM Assay

机译:通过IR-MALDESI MSI和Q Exactive结合在组织中绘制抗逆转录病毒药物的图谱并与LC-MS / MS SRM分析进行比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This work describes the coupling of the IR-MALDESI imaging source with the Q Exactive mass spectrometer. IR-MALDESI MSI was used to elucidate the spatial distribution of several HIV drugs in cervical tissues that had been incubated in either a low or high concentration. Serial sections to those analyzed by IR-MALDESI MSI were homogenized and analyzed by LC-MS/MS to quantify the amount of each drug present in the tissue. By comparing the two techniques, an agreement between the average intensities from the imaging experiment with the absolute quantities for each drug was observed. This correlation between these two techniques serves as a prerequisite to quantitative IR-MALDESI MSI. In addition, a targeted MS2 imaging experiment was also conducted to demonstrate the capabilities of the Q Exactive and to highlight the added selectivity that can be obtained with SRM or MRM imaging experiments.
机译:这项工作描述了IR-MALDESI成像源与Q Exactive质谱仪的耦合。 IR-MALDESI MSI用于阐明在低浓度或高浓度下孵育的宫颈组织中几种HIV药物的空间分布。将通过IR-MALDESI MSI分析的连续切片进行均质化,并通过LC-MS / MS分析,以量化组织中每种药物的含量。通过比较这两种技术,可以观察到成像实验的平均强度与每种药物的绝对量之间的一致性。这两种技术之间的相关性是进行定量IR-MALDESI MSI的前提。此外,还进行了有针对性的MS 2 成像实验,以证明Q Exactive的功能并突出显示可以通过SRM或MRM成像实验获得的附加选择性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号