To measure and compare the contrast to noise ratio (CNR) as a function of dose for the CBCTs produced by the mega‐voltage (MV) imaging beam line (IBL) and the treatment beam line (TBL), and to compare the dose to target and various critical structures of pediatric patients for the IBL CBCT versus standard TBL orthogonal port films.Two Siemens Oncor linear accelerators were modified at our institution such that the MV‐CBCT would operate under an investigational IBL rather than the standard 6MV TBL. Prior to the modification, several CBCTs of an electron density phantom were acquired with the TBL at various dose values. After the modification, another set of CBCTs of the electron density phantom were acquired for various doses using the IBL. The contrast to noise ratio (CNR) for each tissue equivalent insert was calculated. In addition, a dosimetric study of pediatric patients was conducted comparing the 1 cGy IBL CBCT and conventional TBL orthogonal pair port films.The CNR for eight tissue equivalent inserts at five different dose settings for each type of CBCT was measured. The CNR of the muscle insert was 0.8 for a 5 cGy TBL CBCT, 1.1 for a 1.5 cGy IBL CBCT, and 2.8 for a conventional CT. The CNR of the trabecular bone insert was 2.9 for a 5 cGy TBL CBCT, 5.5 for a 1.5 cGy IBL CBCT, and 14.8 for a conventional CT. The IBL CBCT delivered approximately one‐fourth the dose to the target and critical structures of the patients as compared to the TBL orthogonal pair port films.The IBL CBCT improves image quality while simultaneously reducing the dose to the patient as compared to the TBL CBCT. A 1 cGy IBL CBCT, which is used for bony anatomy localization, delivers one‐fourth the dose as compared to conventional ortho‐pair films.PACS number: 87.57.Q, 87.57.cj, 87.53.Jw
机译:测量并比较兆伏(MV)成像束线(IBL)和治疗束线(TBL)产生的CBCT的对比噪声比(CNR)与剂量的关系,并将剂量与针对IBL CBCT与标准TBL正交端口胶片的儿科患者的目标和各种关键结构。我们机构对两台西门子Oncor线性加速器进行了修改,以使MV-CBCT可以在IBL研究性的基础上而不是标准的6MV TBL进行操作。在修改之前,使用不同剂量值的TBL获得了几个电子密度模型的CBCT。修改后,使用IBL获得了不同剂量的另一组电子密度模型的CBCT。计算每个组织等效插入物的对比噪声比(CNR)。此外,还对儿科患者进行了剂量学研究,比较了1 cGy IBL CBCT和常规TBL正交成对端口膜,并针对每种CBCT在五个不同剂量设置下测量了八个组织等效插入物的CNR。对于5 cGy TBL CBCT,肌肉插入物的CNR为0.8,对于1.5 cGy IBL CBCT为1.1,对于常规CT为2.8。对于5 cGy TBL CBCT,小梁骨插入物的CNR为2.9,对于1.5 cGy IBL CBCT为5.5,对于常规CT为14.8。与TBL正交对端口胶片相比,IBL CBCT向患者的目标和关键结构输送的剂量约为四分之一。与TBL CBCT相比,IBL CBCT在提高图像质量的同时减少了对患者的剂量。用于骨解剖定位的1 cGy IBL CBCT的剂量是传统邻位对胶片的四分之一.PACS编号:87.57.Q,87.57.cj,87.53.Jw

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号