首页> 外文OA文献 >Characterization of the Megavoltage Cone-Beam Computed Tomography (MV-CBCT) System on HalcyonTM for IGRT: Image Quality Benchmark, Clinical Performance, and Organ Doses
【2h】

Characterization of the Megavoltage Cone-Beam Computed Tomography (MV-CBCT) System on HalcyonTM for IGRT: Image Quality Benchmark, Clinical Performance, and Organ Doses

机译:IGRT中Halcyntm对兆伏锥梁计算断层扫描(MV-CBCT)系统的表征:图像质量基准,临床表现和器官剂量

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Purpose: The Varian Halcyon includes an ultrafast 6 MV flattening filter free (FFF) cone-beam computed tomography (MV-CBCT). Although a kV-CBCT add-on is available, in the basic configuration MV is used for image guided radiotherapy (IGRT). We characterized the MV-CBCT imager in terms of reproducibility, linearity, field of view (FOV) dependence, detectability of soft-tissue, and the effect of metal implants. The performance of the MV-CBCT in the clinic, including resulting dose to organs, is also discussed herein.Methods: A Gammex phantom was scanned using a Halcyon MV-CBCT and a 120 kVp Siemens Definition Edge CT. Mean and standard deviation of Hounsfield Units (HUs) for different electron density relative to water (ρeW) inserts were extracted. Doses to clinical patients due to MV-CBCT are calculated within Eclipse during treatment planning.Results: A stable and near-linear HU-to-ρeW curve was obtained using the MV-CBCT. As the scan length increased from 10 to 28cm, the linearity of curve improved while the mean HUs decreased by 30%. All soft tissue inserts in the Gammex phantom were distinguishable. A crescent artifact affected HU measurements by up to 40 HUs. Soft-tissue contrast was sufficient for clinical online image-guidance in the low dose (5 MU) mode. Mean doses per fraction to organs-at-risk (OARs) were as high as 6 cGy for head and neck, 5 cGy for breast, and 4 cGy for pelvis patients. Metal rods did not affect HU values or introduce noticeable artifacts.Conclusions: Halcyon's MV-CBCT has sufficient soft tissue contrast for IGRT and lacks metal-induced artifacts. Even though the absolute HU values vary with phantom size and scanning length, the HU-to-ρeW conversions are linear and stable day-to-day. In clinical cases, highest tissue doses from MV-CBCT ranged from 2-7cGy per fraction for various treatment sites, which could be significant for some organs at risk. Dose to out-of-treatment-field organs can be limited by reducing the scan length definition during planning and using the low dose mode. The high quality imaging mode did not provide material advantages over the low dose mode. Adequate IGRT was successfully delivered to multiple tumor sites using MV-CBCT.
机译:目的:瓦里安哈尔西恩包括超快6 MV平坦滤波器的分类(FFF)锥形束计算机断层摄影(MV-CBCT)。虽然千伏-CBCT附加是可用的,在基本配置中MV用于图像引导的放射治疗(IGRT)。我们在再现性,线性度,视场(FOV)的依赖,软组织的可检测和的金属植入物的效果的领域而言,其特征在于所述MV-CBCT成像器。的MV-CBCT在临床,包括所得剂量对器官的性能,还讨论herein.Methods:使用太平MV-CBCT和120的kVp西门子定义边缘CT甲Gammex幻象被扫描。到水(ρeW)插入豪森菲尔德单位(HUS)用于相对于电子密度不同的平均数和标准偏差萃取。剂量临床患者由于MV-CBCT期间治疗planning.Results Eclipse内计算值:接近线性的HU-到ρeW曲线使用MV-CBCT获得稳定的和。随着扫描长度从10增加到28厘米,曲线的线性改善而平均HUS下降30%。在Gammex幻象所有软组织刀片加以区分。一个新月形的神器了40个HUS影响胡测量。软组织对比度足以用于在低剂量(5 MU)模式临床在线图像引导。平均剂量每级分对器官高危(桨)分别高达6 cGy的头部和颈部,5 cGy的乳房,和4 cGy的用于骨盆的患者。金属棒没有影响胡值或引入明显artifacts.Conclusions:太平的MV-CBCT有足够的影像引导放射治疗软组织对比度,并且缺乏金属诱导文物。即使绝对HU值与假想大小和扫描长度各不相同,HU-到ρeW转换是线性和稳定的一天到一天。在临床病例中,最高的组织剂量从MV-CBCT从2-7cGy不等每部分的各种治疗点,这可能是在风险一些器官显著。剂量外的治疗场器官可通过减少规划期间扫描长度的定义和使用低剂量模式的限制。高质量成像模式没有在低剂量模式提供材料的优点。充足的IGRT已成功交付使用MV-CBCT多个肿瘤部位。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号