The amount of weight bearing and the force transmission to the frame have an important influence on the results of treatment with an Ilizarov external fixator. The frame provides beneficial interfragmentary movements and compressive loads at the fracture site through elastic wires. Mobilisation can be achieved by applying a weight-bearing platform at the distal end of the fixator. The effect on the interfragmentary movements and the compressive loads in indirect and direct loading were analysed in this study using a composite tibia bone model. Displacement transducers were attached to measure the interfragmentary movements and to detect relative movements of the bone fragments and movements between the rings. The compressive loads in the osteotomy were measured with loading cells in the defect zone. The weight-bearing platform had a substantial effect on the biomechanical behaviour of the frame. It led to an indirect force transmission through the fixator with respect to the osteotomy, resulting in lower compressive loads, lower interfragmentary movements and higher mechanical stress on the frame.
机译:负重的数量和传递到框架的力对使用Ilizarov外固定器的治疗结果有重要影响。框架通过弹性线在骨折部位提供有益的碎片间运动和压缩载荷。可以通过在固定器的远端使用承重平台来实现动员。在这项研究中,使用复合胫骨模型,分析了在间接和直接负荷中对股间运动和压缩负荷的影响。连接位移传感器以测量碎片间运动并检测骨碎片的相对运动和环之间的运动。用缺损区域的测力计测量截骨术中的压缩负荷。承重平台对车架的生物力学行为有很大影响。它导致通过固定器的间接力相对于截骨术的传递,从而导致较低的压缩载荷,较低的节间运动和较高的机架机械应力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号