A wide variety of existing combinations of target and filter materials (target/filter combinations) are used in mammography equipment. The patient dose depends on the X‐ray quality that is derived from the target/filter combination, and a calibration of the dosimeter that is used in the measurement that corresponds to the specific target/filter combination is necessary. However, dosimeters in mammography are generally calibrated with reference to the X‐ray quality of Mo/Mo or W/Al combinations, and it is unclear whether or not the X‐ray quality that is derived from other target/filter combinations will affect the calibration coefficients. In this paper, the calibration coefficients of different dosimeters were evaluated for target/filter combinations. For an ionization chamber‐type dosimeter, good energy dependence was found and the effect of the target/filter combination was small. However, for a semiconductor dosimeter, a large energy dependence was found, and different calibration coefficients that depended on the target/filter combination used were required.PACS number: 87.59.E‐
机译:乳房X线照相设备使用靶材和过滤材料的多种现有组合(靶材/过滤器组合)。患者的剂量取决于从靶标/滤镜组合获得的X射线质量,因此必须校准用于测量中与特定靶标/滤镜组合相对应的剂量计。但是,乳腺X线摄影剂量计通常参照Mo / Mo或W / Al组合的X射线质量进行校准,尚不清楚从其他目标/滤光片组合得出的X射线质量是否会影响校准系数。在本文中,针对目标/过滤器组合评估了不同剂量计的校准系数。对于电离室型剂量计,发现了良好的能量依赖性,并且目标/过滤器组合的影响很小。但是,对于半导体剂量计,发现能量依赖性很大,并且需要取决于所使用的目标/滤波器组合的不同校准系数.PACS编号:87.59。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号