首页> 美国卫生研究院文献>Cancer Biotherapy Radiopharmaceuticals >Spatial Resolution and Image Qualities of Zr-89 on Siemens Biograph TruePoint PET/CT
【2h】

Spatial Resolution and Image Qualities of Zr-89 on Siemens Biograph TruePoint PET/CT

机译:Zr-89在Siemens Biograph TruePoint PET / CT上的空间分辨率和图像质量

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Purpose: Zirconium-89 (t1/2=78.41 hours) is an ideal metallic radioisotope for immuno–positron emission tomography (PET), given that its physical half-life closely matches the biological half-life of monoclonal antibodies. In this study, the authors measured the spatial resolution and image quality of Zr-89 PET and compared the results against those obtained using F-18 PET, which is widely regarded as the gold standard for comparison of imaging characteristics.>Materials and Methods: The spatial resolution and image qualities of Zr-89 were measured on the Siemens Biograph Truepoint TrueV PET/CT scanner, partly according to NEMA NU2-2007 standards. For spatial resolution measurement, the Zr-89 point source was located at the center of the axial field of view (FOV) and offset 1/4 axial FOV from the center. For image quality measurements, an NEMA IEC Phantom was used. The NEMA IEC Phantom consists of six hot spheres that were filled with Zr-89 solution. Spatial resolution and image quality (%contrast, %background variability [BV], and source to background ratio [SBR]) were assessed to compare the imaging characteristics of F-18 with those of Siemens Biograph Truepoint TrueV.>Results: The transverse and axial spatial resolutions at 1 cm were 4.5 and 4.7 mm for Zr-89, respectively. The %contrast of Zr-89 was 25.5% for the smallest 10 mm sized sphere and 89.8% for the largest 37 mm sized sphere, and for F-18, it was 32.5% for the smallest 10 mm sized sphere and 103.9% for the largest 37 mm sized sphere using the ordered subset expectation maximization (OSEM) reconstruction method. The %BV of F-18 PET was 6.4% for the smallest 10 mm sized sphere and 3.5% for the largest 37 mm sized sphere using the OSEM reconstruction. The SBR of Zr-89 was 1.8 for the smallest 10 mm sized sphere and 3.7 for the largest 37 mm sized sphere, and for F-18, it was 2.0 for the smallest 10 mm sized sphere and 4.1 for the largest 37 mm sized sphere using the OSEM reconstruction method.>Conclusions: This study assessed Zr-89 imaging characteristics using a Siemens Biograph Truepoint TrueV PET/CT scanner and compared the results with those obtained for F-18 PET. Although spatial resolution and image quality of Zr-89 PET were lower compared with F-18 PET, due to longer positron range and low positron branching ratio, Zr-89 is advantageous for immuno-PET due to well-matched half-life with monoclonal antibodies.
机译:>目的:Zirconium-89(t1 / 2 = 78.41小时)是免疫正电子发射断层扫描(PET)的理想金属放射性同位素,因为它的物理半衰期与生物学上的半衰期紧密匹配。单克隆抗体。在这项研究中,作者测量了Zr-89 PET的空间分辨率和图像质量,并将结果与​​使用F-18 PET获得的结果进行了比较,F-18 PET被普遍认为是比较成像特性的金标准。>材料和方法:部分地根据NEMA NU2-2007标准,在Siemens Biograph Truepoint TrueV PET / CT扫描仪上测量Zr-89的空间分辨率和图像质量。对于空间分辨率测量,Zr-89点源位于轴向视场(FOV)的中心,并且偏离中心1/4轴向FOV。对于图像质量测量,使用了NEMA IEC Phantom。 NEMA IEC Phantom由六个充满Zr-89解决方案的热球组成。评估了空间分辨率和图像质量(%对比度,%背景变异性[BV]和源与背景之比[SBR]),以比较F-18的成像特性和Siemens Biograph Truepoint TrueV的成像特性。>结果:< / strong> Zr-89在1 cm处的横向和轴向空间分辨率分别为4.5和4.7 mm。对于最小的10 mm球,Zr-89的对比度百分比为25.5%,对于最大的37 mm球,Zr-89的对比度为89.8%;对于F-18,最小的10 mm球为32.5%,对于F-18,则为103.9%。使用有序子集期望最大化(OSEM)重建方法,最大的37mm大小的球体。使用OSEM重建,最小的10 mm球体的F-18 PET的%BV为6.4%,最大的37 mm球体的BV为3.5%。 Zr-89的SBR对于最小的10BRmm球体为1.8,对于最大的37 mm球体为3.7,对于F-18,对于最小的10 mm球体为2.0,对于最大37 mm球体为4.1。 >结论:该研究使用Siemens Biograph Truepoint TrueV PET / CT扫描仪评估了Zr-89成像特性,并将结果与​​F-18 PET的结果进行了比较。尽管Zr-89 PET的空间分辨率和图像质量比F-18 PET低,但由于正电子作用范围更长且正电子支化比低,Zr-89由于与单克隆抗体的半衰期相匹配,因此对免疫PET有利抗体。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号