Objective: The Siemens Biograph TruePoint TrueV (B-TPTV) positron emission tomography (PET) scanner performs 3D PET reconstruction using a system matrix with point spread function (PSF) modeling (called the True X reconstruction). PET resolution was dramatically improved with the True X method. In this study, we assessed the spatial resolution and image quality on a B-TPTV PET scanner. In addition, we assessed the feasibility of animal imaging with a B-TPTV PET and compared it with a microPET R4 scanner. Methods: Spatial resolution was measured at center and at 8 cm offset from the center in transverse plane with warm background activity. True X, ordered subset expectation maximization (OSEM) without PSF modeling, and filtered back-projection (FBP) reconstruction methods were used. Percent contrast (% contrast) and percent background variability (% BV) were assessed according to NEMA NU2-2007. The recovery coefficient (RC), non-uniformity, spill-over ratio (SOR), and PET imaging of the Micro Deluxe Phantom were assessed to compare image quality of B-TPTV PET with that of the microPET R4. Results: When True X reconstruction was used, spatial resolution was <3.65 mm with warm background activity. % contrast and % BV with True X reconstruction were higher than those with the OSEM reconstruction algorithm without PSF modeling. In addition, the RC with True X reconstruction was higher than that with the FBP method and the OSEM without PSF modeling method on the microPET R4. The non-uniformity with True X reconstruction was higher than that with FBP and OSEM without PSF modeling on microPET R4. SOR with True X reconstruction was better than that with FBP or OSEM without PSF modeling on the microPET R4. Conclusions: This study assessed the performance of the True X reconstruction. Spatial resolution with True X reconstruction was improved by 45 % and its % contrast was significantly improved compared to those with the conventional OSEM without PSF modeling reconstruction algorithm. The noise level was higher than that with the other reconstruction algorithm. Therefore, True X reconstruction should be used with caution when quantifying PET data.
机译:目标:西门子Biograph TruePoint TrueV(B-TPTV)正电子发射断层扫描(PET)扫描仪使用具有点扩散函数(PSF)建模的系统矩阵执行3D PET重建(称为True X重建)。使用True X方法极大地提高了PET的分辨率。在这项研究中,我们评估了B-TPTV PET扫描仪的空间分辨率和图像质量。此外,我们评估了使用B-TPTV PET进行动物成像的可行性,并将其与microPET R4扫描仪进行了比较。方法:在具有温暖背景活动的横向平面中心和距中心偏移8 cm处测量空间分辨率。使用True X,无PSF建模的有序子集期望最大化(OSEM)和滤波反投影(FBP)重建方法。根据NEMA NU2-2007评估对比度百分比(对比度百分比)和背景变异性百分比(百分比BV)。评估了Micro Deluxe Phantom的回收系数(RC),不均匀性,溢出率(SOR)和PET成像,以比较B-TPTV PET和microPET R4的图像质量。结果:当使用True X重建时,空间分辨率为<3.65 mm,背景活动活跃。采用True X重建的对比度百分比和BV百分比高于没有PSF建模的OSEM重建算法的百分比。此外,在microPET R4上使用True X重建的RC高于使用FBP方法和未使用PSF建模方法的OSEM的RC。在microPET R4上没有使用PSF建模的情况下,True X重建的不均匀性高于FBP和OSEM。在microPET R4上不使用PSF建模的情况下,使用True X重建的SOR优于使用FBP或OSEM的SOR。结论:本研究评估了True X重建的性能。与不使用PSF建模重建算法的传统OSEM相比,使用True X重建的空间分辨率提高了45%,并且其对比度百分比显着提高。噪声水平高于其他重建算法。因此,在定量PET数据时,应谨慎使用True X重建。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号