Synthesis and quality control of fluorodeoxyglucose and performance assessment of Siemens MicroFocus 220 small animal PET scanner.

机译:氟脱氧葡萄糖的合成和质量控制以及Siemens MicroFocus 220小动物PET扫描仪的性能评估。

获取原文
获取原文并翻译 | 示例

摘要

The scope of this article is to cover the synthesis and quality control procedures involved in production of Fludeoxyglucose (18F--FDG). The article also describes the cyclotron production of 18F radioisotope and gives a brief overview on operations and working of a fixed energy medical cyclotron.;The quality control procedures for FDG involve radiochemical and radionuclidic purity tests, pH tests, chemical purity tests, sterility tests, endotoxin tests. Each of these procedures were carried out for multiple batches of FDG with a passing rate of 95% among 20 batches.;The article also covers the quality assurance steps for the Siemens MicroPET Focus 220 Scanner using a Jaszczak phantom. We have carried out spatial resolution tests on the scanner, with an average transaxial resolution of 1.775mm with 2-3mm offset. Tests involved detector efficiency, blank scan sinograms and transmission sinograms. A series of radioactivity distribution tests are also carried out on a uniform phantom, denoting the variations in radioactivity and uniformity by using cylindrical ROIs in the transverse region of the final image.;The purpose of these quality control tests is to make sure the manufactured FDG is biocompatible with the human body. Quality assurance tests are carried on PET scanners for efficient performance, and to make sure the quality of images acquired is according to the radioactivity distribution in the subject of interest.
机译:本文的范围涵盖了氟丁氧基葡萄糖(18F--FDG)生产中涉及的合成和质量控制程序。本文还介绍了18F放射性同位素的回旋加速器生产,并简要介绍了固定能量医疗回旋加速器的操作和工作.FDG的质量控制程序包括放射化学和放射性核素纯度测试,pH测试,化学纯度测试,无菌测试,内毒素测试。对20个批次中的多批次FDG以95%的合格率执行了这些程序中的每一个。本文还介绍了使用Jaszczak体模的Siemens MicroPET Focus 220扫描仪的质量保证步骤。我们在扫描仪上进行了空间分辨率测试,平均轴向分辨率为1.775mm,偏移为2-3mm。测试涉及检测器效率,空白扫描正弦图和透射正弦图。还对均匀的模型进行了一系列放射性分布测试,通过在最终图像的横向区域中使用圆柱形ROI来表示放射性和均匀性的变化。这些质量控制测试的目的是确保制造的FDG与人体生物相容。在PET扫描仪上进行质量保证测试以提高性能,并确保所获取图像的质量符合感兴趣对象的放射性分布。

著录项

  • 作者单位

    The University of Texas at San Antonio.;

  • 授予单位 The University of Texas at San Antonio.;
  • 学科 Nuclear physics and radiation.;Biomedical engineering.;Chemistry.
  • 学位 M.S.
  • 年度 2016
  • 页码 81 p.
  • 总页数 81
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号