首页> 美国卫生研究院文献>Archives of Disease in Childhood. Fetal and Neonatal Edition >Accuracy of the volume and pressure displays of high frequency oscillators
【2h】

Accuracy of the volume and pressure displays of high frequency oscillators

机译:高频振荡器的体积和压力显示精度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Objective: To determine the effect of frequency on the accuracy of volume and pressure displays of high frequency oscillators. >Methods: The effect of frequency on the displayed volume of the Stephanie, Dräger Babylog 8000 Plus, and SLE 5000 oscillators was assessed. A sine wave pump delivered a constant tidal volume at frequencies of 5–15 Hz to the patient manifold of the oscillators. The displayed volumes at each frequency were compared with the delivered volume. The effect of frequency on displayed pressure was assessed by connecting the oscillator's patient manifold to a lung model; three types of oscillator were studied (SensorMedics 3100A, SLE 5000, and Stephanie). Airway pressure was measured from the manifold using a pressure transducer and non-compliant tubing; the pressure measuring system had a flat frequency response to 30 Hz. >Results: The SLE 5000 volume display overread the delivered volume (by about 5%), but was not affected by frequency. At 5 Hz, the Dräger Babylog 8000 Plus and the Stephanie underread the delivered volume (by about 20%). Increasing frequency resulted in a greater discrepancy between the delivered and displayed volume with the Stephanie, but a smaller discrepancy with the Dräger Babylog 8000 Plus. Altering frequency had a small effect (maximum difference 6%) on the relation between the displayed and delivered pressure for all three oscillators. >Conclusion: Frequency affects the accuracy of displayed volumes and, to a lesser extent, displayed pressures of high frequency oscillators. The results emphasise that data displayed by new devices should not be uncritically accepted.
机译:>目标:确定频率对高频振荡器的体积和压力显示精度的影响。 >方法:评估了频率对Stephanie,DrägerBabylog 8000 Plus和SLE 5000振荡器显示体积的影响。正弦波泵以5–15 Hz的频率向振荡器的患者总管输送恒定的潮气量。将每个频率下显示的体积与输送的体积进行比较。通过将振荡器的患者总管连接到肺部模型,可以评估频率对显示压力的影响。研究了三种类型的振荡器(SensorMedics 3100A,SLE 5000和Stephanie)。使用压力传感器和非顺应性管道从歧管测量气道压力;压力测量系统对30 Hz的频率响应平坦。 >结果:SLE 5000的体积显示覆盖了已交付的体积(大约5%),但不受频率影响。在5 Hz时,DrägerBabylog 8000 Plus和Stephanie读取的传送音量不足(大约降低20%)。频率增加导致斯蒂芬妮的交付量和显示量之间的差异更大,而德尔格Babylog 8000 Plus的差异更小。改变频率对所有三个振荡器的显示压力和输出压力之间的关系影响很小(最大差异为6%)。 >结论:频率会影响显示的体积的准确性,并在较小程度上影响高频振荡器的显示压力。结果强调,不应不加批判地接受新设备显示的数据。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号