A 58-year-old woman presented with onset of pain associated with a popping sound after a twisting motion 4 years after left total knee revision arthroplasty. She had a complex medical history, including a reported bone cement allergy, and presented to the hospital unable to bear weight. Plain radiographs revealed a broken femoral component, with the femoral metaphyseal sleeve separated from the distal articular component. During surgery, it was observed that the femoral adapter bolt and taper adapter had both fractured. Scanning electron microscopy of the fracture surfaces of the components confirmed that the implant had failed in fatigue, presumably due to high cyclic loads. Failure at this junction has not been described previously. In this type of knee design, we recommend supporting the distal articular component either with bone, augmentation, and/or bone cement to reduce the risk for this mode of failure.
机译:一名58岁的女性在左全膝关节置换术后4年发生扭曲运动后出现疼痛,伴随着突然的爆裂声。她有复杂的病史,包括据报道对骨水泥过敏,并被送往医院无法承受体重。普通X线片显示股骨组件破裂,股骨干phy端套筒与远端关节组件分离。在手术过程中,观察到股骨接合器螺栓和锥形接合器均断裂。部件的断裂表面的扫描电子显微镜证实,植入物疲劳失败,可能是由于高周期性载荷造成的。先前未描述此连接处的故障。在这种类型的膝盖设计中,我们建议用骨,隆起和/或骨水泥支撑远端关节组件,以减少发生这种失败模式的风险。