Comparison of Fourier-Domain and Time-Domain Optical Coherence Tomography in the Measurement of Thinnest Corneal Thickness in Keratoconus

机译:傅里叶域和时域光学相干断层扫描在圆锥角膜最薄角膜厚度测量中的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objective. To compare fourier-domain optical coherence tomography (FD-OCT) and time-domain OCT (TD-OCT) in the determination of thinnest corneal thickness (TCT). Methods. This study included 55 keratoconus patients and 50 healthy volunteers. The RTVue-OCT (FD-OCT) and Visante-OCT (TD-OCT) were used for the measurement of the TCT. Three consecutive scans were performed. The comparison and agreement between the two modalities were analyzed by paired t-test, the Pearson correlation, intraclass correlation coefficient (ICC), and Bland-Altman plots. intraobserver repeatability was analyzed by the intraobserver within-subject standard deviation (S (w)), coefficient of variation (CV(w)), and ICC. Results. The TCT value of normal corneas was higher by RTVue-OCT (530.4 ± 19.7 μm) than Visante-OCT (521.5 ± 18.3 μm) (p < 0.001). For keratoconus eyes, the TCT was 425.0 ± 58.2 μm and 424.4 ± 55.7 μm (difference being 0.6 ± 10.2 μm, p = 0.604). Strong correlation (r = 0.938∼0.985) (ICC = 0.915–0.984) was observed between the two OCTs, and each OCT exhibited excellent repeatability in determining the TCT in all subjects (ICC = 0.984–0.994). Conclusions. The values of TCT obtained from RTVue-OCT and Visante-OCT were highly correlated; however, the two values were different. Both OCT instruments exhibited good intraobserver reliability. The existence of systematic differences suggested that the two instruments cannot be used interchangeably.
机译:目的。为了确定最薄的角膜厚度(TCT),比较傅里叶光学相干断层扫描(FD-OCT)和时域OCT(TD-OCT)。方法。这项研究包括55名圆锥角膜患者和50名健康志愿者。 RTVue-OCT(FD-OCT)和Visante-OCT(TD-OCT)用于测量TCT。进行了三个连续扫描。通过配对t检验,Pearson相关性,类内相关系数(ICC)和Bland-Altman图分析了两种模态之间的比较和一致性。观察者内重复性是通过观察者内受试者标准差(S(w)),变异系数(CV(w))和ICC进行分析的。结果。 RTVue-OCT(530.4±19.7μm)高于Visante-OCT(521.5±18.3μm),正常角膜的TCT值更高(p <0.001)。对于圆锥角膜眼睛,TCT为425.0±58.2μm和424.4±55.7μm(差异为0.6±10.2μm,p = 0.604)。在两个OCT之间观察到强相关性(r = 0.938-0.985)(ICC = 0.915-0.984),并且每个OCT在确定所有受试者的TCT时均表现出极好的可重复性(ICC = 0.984-0.994)。结论。从RTVue-OCT和Visante-OCT获得的TCT值高度相关。但是,这两个值是不同的。两种OCT仪器都具有良好的观察者内部可靠性。系统差异的存在表明这两种工具不能互换使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号