AIMS—To develop a more reproducible method of macular volumetric analysis in order to facilitate serial monitoring of changes in retinal oedema with time.
METHODS—The Heidelberg retinal tomography (HRT) machine was used to scan the macula of 20 normal subjects and 40 diabetics with macular oedema. The volume above reference plane (VARP) within a 2 mm diameter and 3 mm diameter circle was measured twice in each eye. The position of the circle to be measured was carefully defined relative to major retinal vessels. As a modification to the previously published technique, fluctuation of the height of the contour line relative to the focal plane of the eye is compensated for by an adjustment of the reference plane. The position of the circle was controlled relative to major retinal vessels using an acetate. The reproducibility of VARP measurement was assessed by three variables—the standard error of the difference on two isolated VARP measurements, the standard error of the difference between the average of three readings on two different occasions, as well as the coefficient of variation (COV).
RESULTS—Both the 2 mm and 3 mm circles showed good reproducibility in VARP measurements. In normal subjects, the COV of VARP for 2 mm circles and 3 mm circles were 13% and 12% respectively. In patients with macular oedema, the COV of VARP for 2 mm circles and 3 mm circles were 9% and 11% respectively.
CONCLUSION—The modification of VARP measurements between scans of the same eye has improved the COV from 31% to 9% in eyes with diabetic macular oedema.

机译:目的:开发一种可重复性更高的黄斑体积分析方法,以便于随时间连续监测视网膜水肿的变化。
方法:使用Heidelberg视网膜层析成像(HRT)机扫描20名正常受试者的黄斑40例患有黄斑水肿的糖尿病患者。在每只眼睛中两次测量直径2毫米和直径3毫米的圆圈内的参考平面(VARP)上方的体积。相对于主要视网膜血管仔细地确定了要测量的圆的位置。作为对先前公开的技术的修改,轮廓线的高度相对于眼睛的焦平面的波动通过参考平面的调整来补偿。使用乙酸盐相对于主要视网膜血管控制圆圈的位置。通过以下三个变量评估VARP测量的可重复性:两次独立VARP测量的差异的标准误差,两次不同情况下三个读数的平均值之间的差异的标准误差以及变异系数(COV) 。
结果-2 mm和3 mm的圆圈在VARP测量中均显示出良好的可重复性。在正常受试者中,VARP的2 mm圆和3mm圆的COV分别为13%和12%。在黄斑水肿患者中,2毫米眼和3毫米眼的VARP的COV分别为9%和11%。
结论—同一只眼睛两次扫描之间VARP测量值的修改使COV从31%改善患有糖尿病性黄斑水肿的患者中的9%。